Your browser doesn't support javascript.
loading
Design of low-intermediate frequency electrotherapy and pain assessment system / 生物医学工程学杂志
Journal of Biomedical Engineering ; (6): 558-562, 2014.
Article in Chinese | WPRIM | ID: wpr-290716
ABSTRACT
Aiming at the single treatment and the design separation between treatment and assessment in electrotherapy equipment, a kind of system including low-intermediate frequency treatment and efficacy evaluation was developed. With C8051F020 single-chip microcomputer as the core and the circuit design and software programming used, the system realized the random switch of therapeutic parameters, the collection, display and data storage of pressure pain threshold in the assessment. Experiment results showed that the stimulus waveform, current intensity, frequency, duty ratio of the system output were adjustable, accurate and reliable. The obtained pressure pain threshold had a higher accuracy (< 0.3 N) and better stability, guiding the parameter choice in the precise electrical stimulation. It, therefore, provides a reliable technical support for the treatment and curative effect assessment.
Subject(s)
Full text: Available Index: WPRIM (Western Pacific) Main subject: Pain Measurement / Software / Microcomputers / Electric Stimulation Therapy / Equipment Design Language: Chinese Journal: Journal of Biomedical Engineering Year: 2014 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Pain Measurement / Software / Microcomputers / Electric Stimulation Therapy / Equipment Design Language: Chinese Journal: Journal of Biomedical Engineering Year: 2014 Type: Article