The Efficacy of Ultrafiltration after Cardiopulmonary Bypass without Homologous Blood Transfusion for Pediatric Cardiac Surgery / 日本心臓血管外科学会雑誌
Japanese Journal of Cardiovascular Surgery
; : 73-77, 1994.
Article
in Ja
| WPRIM
| ID: wpr-366023
Responsible library:
WPRO
ABSTRACT
Thirty-four patients with congenital cardiac disease were studied to evaluated the role of ultrafiltration after cardiopulmonary bypass without homologous blood transfusion. We used either polypropylene microporous hollow fiber hemoconcentrator (HC-30M or 100M) or polyacrylonitrile microporous hollow fiber hemoconcentrator (PHC-500). Ultrafiltration was useful in the reduction of fluid overloading after cardiopulmonary bypass with extreme hemodilution. Thirty-two patients tolerated the procedure uneventfully without donor blood transfusion and were discharged from the hospital. The values of hematocrit, serum protein and free hemoglobin increased significantly after ultrafiltration with either type of hemoconcentrator. However the degree of concentration of blood components was significantly higher with polyacrylonitrile hemoconcentrator than those with polypropylene hemoconcentrator. These results indicated that ultrafiltration was useful for maintaining water balance after cardiopulmonary bypass without homologous blood transfusion in pediatric cardiac surgery and that polyacrylonitrile microporous hollow fiber hemoconcentrator should be employed in patients with shorter bypass time and less hemolysis.
Full text:
1
Index:
WPRIM
Language:
Ja
Journal:
Japanese Journal of Cardiovascular Surgery
Year:
1994
Type:
Article