Your browser doesn't support javascript.
loading
Calibration application evaluation of self-made fresh blood on Beckman HMX and Abx Pentra 60 / 国际检验医学杂志
International Journal of Laboratory Medicine ; (12): 60-61, 2015.
Article in Chinese | WPRIM | ID: wpr-459238
ABSTRACT
Objective To make the detection results of the different automated hematology analyzers in laboratory to be reliable and comparable.Methods With the Beckman Coulter LH750 after calibration as the reference instrument,the fresh whole blood was extracted from the patient and divided into 3 equal parts after mixing evenly.The first part was determined the target value in the LH750 instrument;the second part was used to calibrate the HMX and ABX PENTRA 60 instruments and the third part served as a calibration validation.After calibration,the fresh whole blood of patient was performed the detection for conducting the accuracy evaluation.Results After calibration,the detection results of RBC,HGB,WBC,HCT,MCV and PLT by each instrument were largely improved and the error of the results was less than 1/3 of CLIA′88 requirements.Conclusion Adopting fresh whole blood for calibrating the hematology analyzer is feasible.

Full text: Available Index: WPRIM (Western Pacific) Type of study: Prognostic study Language: Chinese Journal: International Journal of Laboratory Medicine Year: 2015 Type: Article