Design of new type thermal container for infant / 医疗卫生装备
Chinese Medical Equipment Journal
; (6)1989.
Article
in Zh
| WPRIM
| ID: wpr-590640
Responsible library:
WPRO
ABSTRACT
Objective To develop infant thermal container with forced protection circuit so as to ensure its security. Methods Temperature database was set in thermal container and Cygnal F005 was used as intelligent controller. The state of the sensor was judged by comparison precisely. Double temperature controller was adopted to ensure normal working temperature. When the container went wrong, it could automatically switch to protection circuit to keep the temperature for infant. Results With precise temperature control and steady working station, the new type thermal container for infant could provide a safe and comfortable condition with clean air and fitting temperature and humidity. Conclusion The container is steady, accurate, safe and convenient.
Full text:
1
Index:
WPRIM
Language:
Zh
Journal:
Chinese Medical Equipment Journal
Year:
1989
Type:
Article