Your browser doesn't support javascript.
loading
Combination of High Performance Liquid Chromatographywith Nano-electrospray ionization Chirp Z TransformIon Mobility Spectrometry / 分析化学
Chinese Journal of Analytical Chemistry ; (12): 809-816, 2017.
Article in Chinese | WPRIM | ID: wpr-619965
ABSTRACT
By combining frequency modulation Chirp Z transform ion mobility spectrometers (IMS) and multi nozzle electrospray array ionization source, a method of NanoESI-Chirp Z transform ion mobility spectrometry-high performance liquid chromatography was developed for the determination of n-alkyl ammonium bromide compounds.The parameters of NanoESI-Chirp Z transform IMS such as electric field intensity, solvent composition, and solution flow rate were investigated and optimized.Subsequently, four kinds of n-alkyl ammonium bromide compounds were respectively detected by this developed Chirp Z transform method and Fourier transform method, and the obtained results were compared.The result indicated that the optimum conditions were electric field intensity of 4.5 kV, and ESI solution flow rate of 8 μL/min.Then a test mixture containing tetrabutylammonium bromide, tetrapentylammoniumbromide, tetrahexylammonium bromide, tetraheptylammonium bromide, tetranoctylammoniumbromideandtetrakis(decyl) ammonium bromide was successfully separated and determined by the HPLC-nanoESI-Chirp Z IMS method.Chirp Z transform method provided higher signal to noise ratio compared to conventional signal averaging method, and was superior to FT method in the determination of drift time.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Analytical Chemistry Year: 2017 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Analytical Chemistry Year: 2017 Type: Article