Your browser doesn't support javascript.
loading
Analytical Performance and Reference Interval of a Microfluidic Viscometer, Viscore-300 for the Measurement of Whole Blood Viscosity
Laboratory Medicine Online ; : 1-6, 2018.
Article in Korean | WPRIM | ID: wpr-760480
ABSTRACT

BACKGROUND:

Whole blood viscosity (WBV) refers to the internal resistance that occurs when blood flows through blood vessels. WBV is known to be related to many diseases including cardiovascular and neurovascular diseases. We have investigated the analytical performance and established reference intervals for a newly developed microfluidic viscometer, Viscore-300 (NanoBiz, Korea), used for the measurement of WBV.

METHODS:

We performed a precision test of 240 measurements over 20 days using three control materials. For evaluation of repeatability, a total of 60 WBV measurements were made in 3 whole blood samples 20 times a day. A total of 100 whole blood samples were used to evaluate the accuracy of the Viscore-300 in comparison to a rotating viscometer, DV3T (Brookfield, USA), in accordance with the the Clinical and Laboratory Standards Institute's guidelines. To establish the reference intervals, 122 healthy individuals were enrolled in this study.

RESULTS:

The precision and repeatability results showed that the CV was less than 5% for three samples and two shear rates. In the accuracy test, the mean differences between two viscometers were 0.09 cP (0.9%) and −0.07 cP (−1.4%) at shear rates of 10 s−1 and 300 s−1, respectively. The reference intervals of WBV for men were 6.88–13.52 cP at 10 s−1 and 4.32–6.43 cP at 300 s−1; those of women were 5.74–13.29 cP at 10 s−1 and 3.60–6.12 cP at 300 s−1.

CONCLUSIONS:

Viscore-300 showed excellent precision and accuracy and it might be a good instrument for reporting WBV quickly and accurately.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Blood Vessels / Blood Viscosity / Microfluidics Type of study: Practice guideline Limits: Female / Humans / Male Language: Korean Journal: Laboratory Medicine Online Year: 2018 Type: Article