Study on Electromagnetic Interference Risk of Active Implantable Medical Devices in Multiple Unit Environment / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation
; (6): 302-306, 2020.
Article
in Zh
| WPRIM
| ID: wpr-828200
Responsible library:
WPRO
ABSTRACT
Concerning on the safety risks caused by electromagnetic interference of patients implanted with high-risk active implantable medical devices in the environment of domestic MUs, this study evaluates and focuses on the requirements of electromagnetic compatibility in domestic and international standards for rail transit vehicles, the main mechanism of risks caused by EMI, the actual measurement of environmental data in MUs and the working performance of various active implantables in the compartment. The test results shows that all kinds of active implantable medical device samples works normally in the CRH2A EMU in China, and there is still a large margin between the measured radiation emission in MU and the limit required by the standards.
Key words
Full text:
1
Index:
WPRIM
Main subject:
Pacemaker, Artificial
/
Prostheses and Implants
/
China
/
Electromagnetic Fields
/
Electromagnetic Radiation
Type of study:
Etiology_studies
Limits:
Humans
Country/Region as subject:
Asia
Language:
Zh
Journal:
Chinese Journal of Medical Instrumentation
Year:
2020
Type:
Article