ABSTRACT

Objective:

To evaluate the measurement agreement of Roche 25(OH)D immunoassay(evaluation method) with LC-MS/MS (reference method).

Methods:

A total of 909 residual serum samples from routine health check participants were collected from May to June in 2019. 25(OH)D concentrations were measured by evaluation method and LC-MS/MS, respectively. Passing-bablok regression, intraclass correlation coefficient (ICC), Bland Altman plots and Kappa test were used to analyze the consistency and bias on the results derived from the two measurement methods.

Results:

The 25(OH)D concentration derived from evaluation method was significantly different from those from LC-MS/MS method ( P<0.001). Slope of regression for evaluation method and LC-MS/MS was 0.962(95% CI 0.919-1.007), while intercept was -0.185 (95% CI -1.191-0.745). The ICC was 0.765 (95% CI 0.735-0.792). Altman plot showed that the average deviation between evaluation method and LC-MS/MS was -0.902 ng/ml (0.300%). The coincidence rate of evaluation method′s judgment of vitamin D sufficiency, insufficiency and deficiency with LC-MS/MS was 83.39%, and the weighted Kappa values was 0.790.

Conclusion:

Roche automatic 25(OH)D immunoassay shows acceptable correlation and agreement with LC-MS/MS, however, it is to note that the deviation between immunoassay and LC-MS/MS may lead to wrong judgment of vitamin D nutritional status. It is recommended that each laboratory should establish own corresponding reference values for 25(OH)D concentrations derived from these two methods.