Design of Treatment System for Poor Peripheral Circulation Using Air Wave Pressure Based on STM32 Microprocessor / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 42-46, 2020.
Article in Chinese | WPRIM | ID: wpr-942694
ABSTRACT
This study designs an intermittent pneumatic pressurization device with STM32 series single chip as the core. The working state of the air pump and the plurality of air chambers is controlled by the IO port of the single chip microcomputer, and the circulating inflation of the plurality of air bags is realized. The pressure monitoring system consists of a silicon piezoresistive pressure sensor, a high-precision operational amplifier and a 12-bit AD converter which monitors the gas pressure of each gas path in real time to ensure the safety of the equipment. The system is easy to operate, simple in function, and has strong practicability.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Pressure / Blood Circulation / Microcomputers / Cardiovascular System / Equipment Design Limits: Humans Language: Chinese Journal: Chinese Journal of Medical Instrumentation Year: 2020 Type: Article