Acceptance test of PET/CT based on national standard and the NEMA measurement program / 中国医学装备
China Medical Equipment ; (12): 16-22, 2024.
Article en Zh | WPRIM | ID: wpr-1026478
Biblioteca responsable: WPRO
ABSTRACT
Objective:To perform acceptance test and performance assessment for Siemens Biograph Vision 600 positron-emission tomography/computed tomography(PET/CT)according to the national health industry standard WS 817-2023.Methods:Spatial resolution,sensitivity,scatter fraction,count loss and random coincidence,correction accuracy of count loss and random coincidence,time-of-flight(TOF)resolution of the PET component within the PET/CT system were tested through the measurement program(NU2-2018)of National Electrical Manufacturers Association(NEMA),which was installed inside of the equipment,in accordance with the requirement of national health industry standard WS 817-2023.The PET/CT registration accuracy was measured through Gantry_offset acquisition program that was built into the equipment.Results:The transversely and axially spatial resolutions of Biograph Vision 600 PET/CT were respectively 3.69 mm and 4.10 mm at 1 cm away from the center of visual field,and were respectively 4.26 mm and 4.89 mm at 10 cm away from the center of visual field,and were respectively 4.68 mm and 4.89 mm at 20 cm away from the center of visual field.The sensitivity of 10 cm away from center and radial of visual field were respectively 16.12 kcps/MBq and 16.00 kcps/MBq.The peak value of noise equivalent count rate(NECR)was 281.60 kcps,and the corresponding radioactivity concentration of peak value was 30.69 kBq/ml.The NECR peak value,scatter fraction and maximum value of the error of relative count rate were respectively 38.17% and 4.0%.The TOF resolution was 209.87 ps when the radioactivity concentration was 5.3 kBq/mL.The registration accuracy values of Biograph Vision 600 PET/CT were 0.347 mm,-0.226 mm and 3.659 mm at the directions of x,y and z axis.Conclusion:It is feasible to perform the acceptance test according to the WS 817-2023 standard through uses the NEMA NU2-2018 standard measurement program that is installed inside of the equipment.The performance indicators can meet requirement of standard as the current national standard GB/T 18988.1-2013 and the health industry standard WS 817-2023 that will being implemented in the test of Biograph Vision 600 PET/CT,which can pass acceptance.
Palabras clave
Texto completo: 1 Índice: WPRIM Idioma: Zh Revista: China Medical Equipment Año: 2024 Tipo del documento: Article
Texto completo: 1 Índice: WPRIM Idioma: Zh Revista: China Medical Equipment Año: 2024 Tipo del documento: Article