Performance evaluation of AU2700 biochemical analyzer for detecting serum glycated albumin / 国际检验医学杂志
Article en Zh | WPRIM | ID: wpr-461470
Biblioteca responsable: WPRO
ABSTRACT
Objective To use the AU2700 biochemical analyzer to evaluate the performance of the reagent kit for determining se-rum glycated albumin(GA)by the enzymatic method.Methods On the basis of the relevant documents of CLSI-EP,the perform-ance evaluation of the Lucica GA-L reagent kit for detecting GA was performed on the precision,accuracy,linearity range and refer-ence range;the correlation between the results detected by the two kinds of reagent kit (enzymatic method)was analyzed.Results The reagent had high precision(CV<3%);the dilution had good linearity;the average recovery rate of GA was 97.7%,the average recovery rate of albumin(ALB)was 96.43%;the detection results by the two kinds of reagent kit had good correlation (r2 =0.976, P <0.01).Conclusion The Lucica GA-L reagent kit shows the good analytical performance on the AU2700 biochemical analyzer, which is consistent to the manufacturers statement and can be better used in clinic.