Magnetic field numerical calculation and analysis for magnetic coupling of centrifugal blood pump for extracorporeal circulation / 生物医学工程学杂志
Journal of Biomedical Engineering ; (6): 1244-1249, 2013.
Article Dans Zh | WPRIM | ID: wpr-259731
Responsable en Bibliothèque : WPRO
ABSTRACT
This paper mainly studies the driving system of centrifugal blood pump for extracorporeal circulation, with the core being disc magnetic coupling. Structure parameters of disc magnetic coupling are related to the ability of transferring magnetic torque. Therefore, it is necessary to carry out disc magnetic coupling permanent magnet pole number (n), air gap length (L(g)), permanent magnet thickness (L(m)), permanent magnet body inside diameter (R(i)) and outside diameter (R(o)), etc. thoroughly. This paper adopts the three-dimensional static magnetic field edge element method of Ansys for numerical calculation, and analyses the relations of magnetic coupling each parameter to transmission magnetic torque. It provides a good theory basis and calculation method for further optimization of the disc magnetic coupling.
Sujets)
Texte intégral: 1 Indice: WPRIM Sujet Principal: Moment de torsion / Circulation extracorporelle / Champs magnétiques / Magnétisme langue: Zh Texte intégral: Journal of Biomedical Engineering Année: 2013 Type: Article