Influence of projection data correction on digital breast tomosynthesis imaging / 南方医科大学学报
Article de Zh | WPRIM | ID: wpr-273766
Bibliothèque responsable: WPRO
ABSTRACT
<p><b>OBJECTIVE</b>To investigate the effect of detector performance during digital breast tomography (DBT) projection data acquisition on reconstructed image quality.</p><p><b>METHODS</b>With reference to the traditional detector data correction method and the specific data acquisition pattern in DBT imaging, we utilized dark field correction, light field and its gain correction for processing the projection data collected by the detector. The reconstructed images were evaluated using iterative reconstruction method based on total generalized variation (TGV).</p><p><b>RESULTS</b>In physical breast phantom experiment, the proposed method resulted in a reduced Heel effect caused by nonuniform photon number. The reconstructed DBT images after correction showed obviously improved image quality especially in the details with a low contrast.</p><p><b>CONCLUSION</b>The dark field correction, light field and its gain correction process for DBT image reconstruction can improve the image quality.</p>
Texte intégral: 1 Indice: WPRIM langue: Zh Texte intégral: Journal of Southern Medical University Année: 2017 Type: Article