Interruption of bispectral index monitoring by nerve integrity monitoring during tympanoplasty: A case report / 대한마취과학회지
Korean Journal of Anesthesiology
; : 161-163, 2013.
Article
de En
| WPRIM
| ID: wpr-59806
Bibliothèque responsable:
WPRO
ABSTRACT
We report that intraoperative NIM-2 monitoring devices can interfere with bispectral index monitoring. A 45-year-old male with chronic otits media underwent tympanolasty under general anesthesia with NIM-2 monitoring and bispectral index monitoring at our institution. And then, bispectral index monitoring was severely interrupted by facial nerve monitoring.
Mots clés
Texte intégral:
1
Indice:
WPRIM
Sujet Principal:
Surveillance peropératoire
/
Nerf facial
/
Moniteurs d'évaluation de la conscience
/
Anesthésie générale
Limites du sujet:
Humans
/
Male
langue:
En
Texte intégral:
Korean Journal of Anesthesiology
Année:
2013
Type:
Article