Your browser doesn't support javascript.
loading
Analytical Performance of New ARCHITECT AFP Assay: Comparison with the Current Assay
Article Dans En | WPRIM | ID: wpr-713886
Responsable en Bibliothèque : WPRO
ABSTRACT
Alpha-fetoprotein (AFP) is frequently used for hepatocellular carcinoma (HCC) diagnosis and surveillance. Although the current ARCHITECT AFP (List number 7K67) assay range is 0–350 ng/mL, all samples with test results between 200 and 350 ng/mL must be diluted and retested until their levels are <200 ng/mL. A new ARCHITECT AFP (8100/3P36) assay with a dynamic range of up to 2000 ng/mL has been introduced. The aim of this study was to perform a method comparison between the current ARCHITECT AFP assay and the new assay. The precision study showed excellent results for both high and low controls. There was a positive correlation between the two assay systems and clinical samples. The new ARCHITECT AFP assay with a wide assay range demonstrated good analytical performance. Therefore, the current ARCHITECT AFP assay could be replaced by the new assay, which is more convenient and minimizes manual labor.
Sujets)

Texte intégral: 1 Indice: WPRIM Sujet Principal: Alphafoetoprotéines / Carcinome hépatocellulaire / Diagnostic / Méthodes Type d'étude: Diagnostic_studies / Guideline langue: En Texte intégral: Laboratory Medicine Online Année: 2018 Type: Article