Your browser doesn't support javascript.
loading
Rapid determination of acetaminophen in plasma by LC-MS/MS / 中华劳动卫生职业病杂志
Article Dans Zh | WPRIM | ID: wpr-986014
Responsable en Bibliothèque : WPRO
ABSTRACT

Objective:

To establish a method for the rapid determination of acetaminophen (APAP) in human plasma by LC-MS/MS.

Methods:

The plasma samples were extracted by methanol and acetonitrile (1 1) and purified directly. C(18) column was used for sample separation. The mobile phase were methanol (5 mmol/L ammonium acetate) and water (5 mmol/L ammonium acetate). Samples were analyzed by LC MS/MS with the electrospray ionization multi reaction monitoring (MRM) mode.

Results:

The calibration curves of APAP was linear in the concentration range of 0~10 mg/L, the correlation coefficient (r) was greater than 0.999 0. The relative standard deviation within and between batches was less than 10%. The recovery rate were 96.81%~101.7%. The detection limit of the method was 0.1 μg/L and the lower limit of quantification was 0.3 μg/L.

Conclusion:

This method has strong specificity, high sensitivity and reliable determination results. It is suitable for the rapid analysis of clinical plasma samples.
Sujets)

Texte intégral: 1 Indice: WPRIM Sujet Principal: Chromatographie en phase liquide à haute performance / Chromatographie en phase liquide / Méthanol / Spectrométrie de masse en tandem / Acétaminophène Limites du sujet: Humans langue: Zh Texte intégral: Chinese Journal of Industrial Hygiene and Occupational Diseases Année: 2023 Type: Article
Texte intégral: 1 Indice: WPRIM Sujet Principal: Chromatographie en phase liquide à haute performance / Chromatographie en phase liquide / Méthanol / Spectrométrie de masse en tandem / Acétaminophène Limites du sujet: Humans langue: Zh Texte intégral: Chinese Journal of Industrial Hygiene and Occupational Diseases Année: 2023 Type: Article