Avaliação da captura de fragmentos por meio da filtração intra-aórtica em pacientes submetidos à troca valvar aórtica / Particulate emboli capture by an intra-aortic filter device during aortic valve replacement
Rev. bras. cir. cardiovasc
; Rev. bras. cir. cardiovasc;23(3): 431-435, jul.-set. 2008. ilus, tab
Article
em En, Pt
| LILACS
| ID: lil-500535
Biblioteca responsável:
BR1.1
RESUMO
OBJETIVO:
O objetivo deste estudo foi avaliar a atividade embólica de pacientes portadores de estenose aórtica calcificada submetidos a troca valvar aórtica por meio da filtração intra-aórtica com dispositivo EMBOL-X® System (Edwards Lifesciences Inc., Mountain View, CA, USA).MÉTODOS:
De janeiro de 2007 a julho de 2007, foi utilizado o filtro intra-aórtico EMBOL-X após o despinçamento aórtico em 13 portadores de estenose aórtica calcificada submetidos a troca valvar aórtica consecutivamente. A média de idade dos pacientes foi 63,7 anos (34-79) e 61,5% eram do sexo feminino. A média do tempo de CEC foi 60,2±7,5 (45-72) minutos e a média do tempo despinçamento aórtico foi 50±7,5 (35-63) minutos. Após a retirada dos filtros, eles foram fixados em formalina, analisados macroscopicamente e quantificados os fragmentos capturados. Foi realizado exame histológico do material capturado. RESULTADOS Não foi observado nenhum caso de complicação neurológica. Nenhum paciente apresentou insuficiência renal pós-operatória. Não houve óbitos hospitalares. Partículas embólicas foram encontradas em cinco (38,5%) dos filtros. Das partículas embólicas capturadas, em dois (40%) filtros havia fibrina, dois (40%) apresentavam tecido conjuntivo, um (20%) continha hemácias e em um não foi possível determinar a natureza.CONCLUSÃO:
O dispositivo EMBOL-X® System foi efetivo na captação de fragmentos intra-aórticos na substituição da valva aórtica em pacientes com estenose aórtica calcificada.ABSTRACT
OBJECTIVE:
This study aims to analyze the embolic activity in patients with calcified aortic stenosis who underwent aortic valve replacement using intra-aortic filtration with an EMBOL-X® System device (Edwards Lifesciences Inc., Mountain View, CA, USA).METHODS:
From January 2007 to July 2007, 13 consecutive patients with calcified aortic stenosis, who underwent isolated aortic valve replacement using intra-aortic filtration by an EMBOL-X® System for 5 minutes after aortic clamp release, were evaluated. Mean patient age was 63.7 years (range 34 to 79 years) and 61.5% were female. The mean bypass time was 60.2 ± 7.5 minutes (range 45 to 72 minutes) and the mean cross-clamp time was 50 ± 7.5 minutes (range 35 to 63 minutes). Following removal, each filter was fixed in formalin and analyzed macroscopically with the captured fragments being counted. Histological examinations of the captured material were performed.RESULTS:
There were no strokes or gross neurological events. There were no cases of postoperative renal failure. No deaths were reported during hospitalization. Particulate emboli were found in five (38.5%) of the filters. On histological analysis of the particulate emboli captured, two (40%) contained fibrin, two (40%) presented conjunctive tissue, one (20%) contained red blood cells and in one it was not possible to determine the nature of the particulates captured.CONCLUSION:
The EMBOL-X® System device was effective in particulate emboli capture in aortic valve replacement surgery of patients with calcified aortic stenosis.Palavras-chave
Texto completo:
1
Índice:
LILACS
Assunto principal:
Estenose da Valva Aórtica
/
Calcinose
/
Implante de Prótese de Valva Cardíaca
/
Embolia
/
Filtração
Limite:
Adult
/
Aged
/
Female
/
Humans
/
Male
Idioma:
En
/
Pt
Revista:
Rev. bras. cir. cardiovasc
Assunto da revista:
CARDIOLOGIA
/
CIRURGIA GERAL
Ano de publicação:
2008
Tipo de documento:
Article
/
Congress and conference