Validação laboratorial de um método automatizado de dosagem da atividade de adenosina desaminase em líquido pleural e em líquido cefalorraquidiano / Laboratorial validation of an automated assay for the determination of adenosine deaminase activity in pleural fluid and cerebrospinal fluid
J. bras. pneumol
; J. bras. pneumol;34(12): 1033-1039, dez. 2008. ilus, tab
Article
em En, Pt
| LILACS
| ID: lil-503816
Biblioteca responsável:
BR1.1
RESUMO
OBJETIVO:
A incidência global de tuberculose reforça a necessidade de melhores ensaios para o diagnóstico desta doença, principalmente da tuberculose extrapulmonar. O objetivo do trabalho foi validar o desempenho de um método automatizado para a determinação da atividade de adenosina desaminase (ADA) no líquido pleural (LP) e no líquido cefalorraquidiano (LCR), comparando-o com um método convencional (Giusti modificado).MÉTODOS:
Selecionaram-se 134 amostras da rotina laboratorial 94 de LP e 40 de LCR. Foram realizadas as determinações da atividade de ADA através dos dois métodos. Calculou-se a precisão inter- e intra-ensaios, análise de regressão linear, testes de concordância simples e médias das diferenças.RESULTADOS:
Os coeficientes de correlação para as amostras de LP e LCR foram, respectivamente, 0,96 e 0,95. A precisão interensaio foi determinada pela média de 21 amostras replicadas em ensaios diferentes para 3 níveis de atividade baixa, média e alta. Os coeficientes de variação em porcentagem ( por centoCV) foram, respectivamente, 5,9, 8,1 e 5,8 para amostras de LP; e 21,9, 18,6 e 13,8 para amostras de LCR, respectivamente. A precisão intra-ensaio em por centoCV foi, respectivamente, 1,3 e 11,7 por cento para amostras de LP e LCR. A concordância entre os dois métodos em amostras de LP e LCR foi, respectivamente, 96,8 por cento e 100 por cento, considerando-se como valores de referência para o diagnóstico de TB 40 U/L (convencional) e 30 U/L (automatizado) em amostras de LP, e 9 U/L em amostras de LCR para os dois métodos.CONCLUSÕES:
Os resultados validaram o método automatizado de determinação da atividade de ADA para o uso em amostras de LP e LCR como alternativa ao método convencional.ABSTRACT
OBJECTIVE:
The incidence of tuberculosis worldwide has emphasized the need for better assays designed to diagnose the disease, principally the extrapulmonary form. The objective of the present study was to validate the performance of an automated method for the determination of adenosine deaminase (ADA) activity in pleural fluid (PF) and cerebrospinal fluid (CSF), comparing it with a conventional method (the modified Giusti method).METHODS:
In total, 134 samples were selected from among those tested in our laboratory 94 PF samples and 40 CSF samples. The ADA activity was determined using the two methods. Inter- and intra-assay precision was determined, linear regression analysis was performed, simple concordance tests were conducted, and the means of the differences were calculated.RESULTS:
The correlation coefficients for PF and CSF samples were, respectively, 0.96 and 0.95. Inter-assay precision was determined using 21 replicates at 3 different activity levels low, medium and high. The percentage coefficient of variation ( percentCV) was, respectively, 5.9, 8.1 and 5.8 for PF samples, compared with 21.9, 18.6 and 13.8 for CSF samples. Intra-assay precision in percentCV was 1.3 and 11.7, respectively, for PF and CSF samples. The concordance between the methods in PF and CRF samples was, respectively, 96.8 percent and 100 percent, considering the reference values for the diagnosis of TB to be 40 U/L (conventional) and 30 U/L (automated) in PF samples, versus 9 U/L (for both methods) in CSF samples.CONCLUSIONS:
The results validate the use of the automated method of determining ADA activity in PF and CSF samples as an alternative to the conventional method.Palavras-chave
Texto completo:
1
Índice:
LILACS
Assunto principal:
Derrame Pleural
/
Kit de Reagentes para Diagnóstico
/
Tuberculose Pulmonar
/
Adenosina Desaminase
/
Ensaios Enzimáticos Clínicos
Tipo de estudo:
Prognostic_studies
Limite:
Humans
Idioma:
En
/
Pt
Revista:
J. bras. pneumol
Assunto da revista:
PNEUMOLOGIA
Ano de publicação:
2008
Tipo de documento:
Article