Dosimetria de pacientes e médicos em intervenções coronárias percutâneas em Recife, Pernambuco, Brasil / Radiation dosimetry in patients and physicians during percutaneous coronary angioplasty in Recife, Pernambuco, Brazil
Radiol. bras
;
44(2): 90-96, mar.-abr. 2011. ilus, tab
Artigo
em Português
| LILACS
| ID: lil-588194
RESUMO
OBJETIVO:
Este trabalho teve como objetivo estimar os valores de doses de radiação recebidas por médicos e pacientes em procedimentos intervencionistas cardíacos realizados em um hospital público na cidade de Recife, Pernambuco.MATERIAIS E MÉTODOS:
As medidas foram determinadas em 31 pacientes adultos, dos quais 22 tiveram acompanhamento clínico após o procedimento, e em dois cardiologistas com mais de dez anos de experiência. Parâmetros de irradiação para cada procedimento foram registrados.RESULTADOS:
Os valores obtidos para a dose absorvida máxima na pele do paciente variaram entre 612 e 8.642 mGy, sendo que 53 por cento foram maiores que 2.000 mGy, valores estes que podem causar efeitos determinísticos. Com relação aos médicos, a dose efetiva média por procedimento foi de 11 µSv e os valores médios do equivalente de dose nas extremidades, mais altos, foram 923 µSv no pé esquerdo, 514 µSv no pé direito, 382 µSv na mão esquerda e 150 µSv no olho esquerdo. Dependendo do número de procedimentos, as doses recebidas pelos médicos podem exceder os valores limites de doses estabelecidos pelas normas nacionais e internacionais.CONCLUSÃO:
Os resultados obtidos sinalizam a necessidade de adoção de estratégias para otimização da proteção radiológica tanto de pacientes quanto de médicos.ABSTRACT
OBJECTIVE:
The present study was aimed at estimating the values of radiation doses received by physicians and patients during interventional cardiology procedures performed at a public hospital in the city of Recife, Pernambuco, Brazil. MATERIALS ANDMETHODS:
Measurements were made in two cardiologists with more than ten years of experience and in 31 adult patients, with 22 of them being clinically followed-up after the procedure. The individual irradiation parameters were documented.RESULTS:
The values for maximum patients skin dose ranged between 612 and 8,642 mGy, achieving more than 2,000 mGy in 53 percent of the patients; such dose values may cause deterministic effects. As regards the physicians, the mean effective dose per procedure was 11 µSv, and the highest mean equivalent doses in the limbs were 923 µSv in the left foot, 514 µSv in the right foot, 382 µSv in the left hand, and 150 µSv in the left eye. Depending on the number of procedures, the doses received by the physician may exceed the dose limits established by the Brazilian and international standards.CONCLUSION:
The obtained results indicate the necessity of adopting strategies for optimizing the radiological protection for both patients and physicians.
Texto completo:
DisponíveL
Índice:
LILACS (Américas)
Tipo de estudo:
Guia de Prática Clínica
País/Região como assunto:
América do Sul
/
Brasil
Idioma:
Português
Revista:
Radiol. bras
Assunto da revista:
Diagnóstico por Imagem
/
Radiologia
Ano de publicação:
2011
Tipo de documento:
Artigo
País de afiliação:
Brasil
Instituição/País de afiliação:
Hospital da Aeronáutica de Pernambuco/BR
/
Hospital das Clínicas de Pernambuco/BR
/
Instituto Federal de Ciências, Educação e Tecnologia de Pernambuco/BR
/
Instituto de Medicina Integral Professor Fernando Figueira/BR
/
Universidade Federal de Pernambuco/BR
Similares
MEDLINE
...
LILACS
LIS