Monoclonal anti-CD47 interference in pre-transfusion test and mitigation measures, a case report / 中国输血杂志
Chinese Journal of Blood Transfusion
; (12): 290-292, 2021.
Article
em Zh
| WPRIM
| ID: wpr-1004567
Biblioteca responsável:
WPRO
ABSTRACT
【Objective】 To discuss the interference of anti-CD47 in pre-transfusion test and the mitigation measures. 【Methods】 Blood sample of one patient received anti-CD47 treatment was collected to conduct routine serological tests including ABO/Rh phenotype, direct anti-human globulin test, irregular antibody screening, antibody identification and cross-match. Packed platelet from multiple type O blood donors was used to absorb with patient′s plasma. The patient′s plasma was absorbed with CCDee, ccDEE and ccdee red cells, respectively. Anti-IgG monoclonal Gamma-clone which lacks reactivity with human subclass IgG4 was used to perform antibody screening and cross-match. Capture-R was used to perform antibody screening. 【Results】 The direct anti-human globulin test was positive(1+ ), the reactivity in all phases was strong positive(3+ -4+ ). The anti-CD47 was eliminated after platelet and red cells absorption. Antibody screening became negative using Gamma-clone and Capture-R, and cross-match successfully using Gamma-clone. 【Conclusion】 Anti-CD47 monoclonal antibody can interfere with pre-transfusion test and cross matching. To remove the interference of anti-CD47 requires the use of Gamma-clone anti-IgG in the indirect antiglobulin testing or Capture-R.