Operation principle and failure case analysis of Gambro Prismaflex continuous renal replacement therapy equipment / 中国医学装备
China Medical Equipment
; (12): 178-181, 2024.
Article
em Zh
| WPRIM
| ID: wpr-1026549
Biblioteca responsável:
WPRO
ABSTRACT
By analyzing the design principles and treatment modes of Gambro Prismaflex CRRT device,based on the basic structure and treatment process,and the typical failure cases of pressure joints and scale zeroing test failures in continuous renal replacement therapy(CRRT)equipment were analyzed,the targeted solution and maintenance strategies were proposed to ensure the stable and efficient operation of CRRT equipment.
Texto completo:
1
Índice:
WPRIM
Idioma:
Zh
Revista:
China Medical Equipment
Ano de publicação:
2024
Tipo de documento:
Article