Objective To construct an immunosensor for detecting CD4+ T lymphocytes without labeling .Methods The staphy-lococcus protein A(SPA) method was adopted to conduct the oriented immobilization of CD4 monoclonal antibodies on the gold in-terdigitated microelectrode surface for capturing CD4+ T lymphocytes .Then cyclic voltammetry(CV) method was used to conduct the representation of modification situation on the gold interdigitated microelectrode surface .Finally the electrochemical impedance spectroscopy(EIS) was used to detect the impedance of CD4+ T lymphocytes captured by the immunosensor .The standard curve was drawn by the impedance values change obtained by the equivalent electric circuit fitting .Results The linear range of this im-munosensor for detecting CD4+ T lymphocytes was (5 × 103 -5 .0 × 106 )/mL ,with lower detection limit of 5 .0 × 102/mL .Conclu-sion The constructed immunosensor has accurate and reliable detection results uhidn is simple to operate accurate ,convenient and cheap ,which might be expected to be used in the real-time detection system ,and offers help for realizing rapid ,accurate and inex-pensive CD4+ T lymphocyte count .
Palavras-chave
Texto completo: 1 Índice: WPRIM Tipo de estudo: Diagnostic_studies Idioma: Zh Revista: International Journal of Laboratory Medicine Ano de publicação: 2017 Tipo de documento: Article