Comparison of efficacy and safety of two different parameters of transscleral cyclophotocoagulation / 国际眼科杂志(Guoji Yanke Zazhi)
International Eye Science
; (12): 1771-1774, 2019.
Article
em Zh
| WPRIM
| ID: wpr-750501
Biblioteca responsável:
WPRO
ABSTRACT
@#AIM:To compare the efficacy of transscleral cyclophotocoagulation(TSCP)with two different laser parameters.<p>METHODS: A total of 46 cases with refractory glaucoma(46 eyes)who underwent TSCP from January 2014 to December 2018 were analyzed retrospectively. According to the laser parameters, all cases were divided into low-power group(treated by low-power TSCP)and conventional parameter group(treated by conventional parameter TSCP). Intraocular pressures(IOP), best-corrected visual acuity(BCVA), number of anti-glaucoma drugs and complications were observed and compared between the two groups.<p>RESULTS: There was no significant difference of IOP between the two groups at postoperative 1d, 1wk, 1mo and 3mo(all <i>P</i>>0.05). The total success rate was 87% in the low-power group while 83% in the conventional parameter group(<i>P</i>=1.000).The overall incidence of complications of the low-power group(39%)was lower than that of the conventional parameter group(78%)(<i>P</i>=0.007)at postoperative 3mo. The early postoperative pain score of the low-power group was significantly lower than that of the conventional parameter group(<i>P</i>=0.007).<p>CONCLUSION:Compared with conventional parameters, TSCP with lower power may reduce IOP similarly with less pain and fewer complications.
Texto completo:
1
Índice:
WPRIM
Idioma:
Zh
Revista:
International Eye Science
Ano de publicação:
2019
Tipo de documento:
Article