Your browser doesn't support javascript.
loading
Validación del método analítico por cromatografía líquida de alta resolución para la cuantificación de conservantes en jugo de naranja / Validation of the analytical method by liquid chromatography of high resolution for the quantification of preservatives in orange juice
Trigo Orsini, Myriam Lina; Aban Aramayo, Elba Amalia.
Affiliation
  • Trigo Orsini, Myriam Lina; Universidad Mayor de San Andrés. Laboratorio de Diagnostico e Investigación en Salud (SELADIS). La Paz. BO
  • Aban Aramayo, Elba Amalia; Universidad Mayor de San Andrés. Laboratorio de Diagnostico e Investigación en Salud (SELADIS). La Paz. BO
Con-ciencia (La Paz) ; 8(2): 61-74, 2020. ilus., tab.
Article в Es | LILACS, LIBOCS | ID: biblio-1147992
Ответственная библиотека: BO138.1
RESUMEN

INTRODUCCIÓN:

la validación es el establecimiento de la evidencia documental de que un procedimiento analítico conducirá, con un alto grado de seguridad a la obtención de resultados precisos y exactos dentro de las especificaciones y los atributos de calidad previamente establecidos.

OBJETIVO:

Validar el método analítico por cromatografía líquida de alta resolución para la cuantificación de conservantes en jugo de naranja.

MÉTODOS:

el método fue validado siguiendo los parámetros referidos en la USP 40 ­ NF 35 especificidad, linealidad, precisión (repetibilidad instrumental, repetibilidad del método y precisión intermedia) y exactitud.

RESULTADOS:

el método analítico propuesto es selectivo/específico, presenta una excelente linealidad dentro del rango de 100 a 1000 ppm, la Precisión presentó resultados de CV < 2% siendo óptimos. La recuperación se encuentra entre 97 a 102% demostrando excelente exactitud.

CONCLUSIÓN:

se cumplieron todos los parámetros y criterios de validación por lo tanto el método es específico, selectivo, lineal en el intervalo de concentraciones de 100 a 1000 ppm, sensible, preciso, reproducible y exacto
ABSTRACT

INTRODUCTION:

validation is the establishment of documentary evidence that an analytical procedure will lead, with a high degree of security, to obtaining precise and exact results within the specifications and previously established quality attributes.

OBJECTIVE:

validate the analytical method by high resolution liquid chromatography for the quantification of preservatives in orange juice.

METHODS:

the method was validated following the parameters referred to in USP 40 - NF 35, they are specificity, linearity, precision (instrumental repeatability, repeatability of the method and intermediate precision) and accuracy.

RESULTS:

the proposed analytical method is selective/specific, it presents an excellent linearity within the range of 100 to 1000 ppm, the precision presented results of CV <2% being optimal. The recovery is between 97 to 102%

CONCLUSION:

all the parameters and validation criteria were met, therefore the method is specific, selective, linear in the concentration range of 100 to 1000 ppm, sensitive, precise, reproducible and exact