Your browser doesn't support javascript.
loading
Feasibility and safety of a new portable endoscopic system for the diagnosis and treatment of abdominal trauma in animal models / 中华消化内镜杂志
Article в Zh | WPRIM | ID: wpr-1029589
Ответственная библиотека: WPRO
ABSTRACT

Objective:

To compare the feasibility and safety of a new portable endoscopic system and the conventional endoscopic system for the detection and emergency treatment of abdominal trauma in animal models.

Methods:

Three healthy Bama pigs, which were fasted and water deprivation for 8 h before surgery and then underwent induction anesthesia. A layer-by-layer incision was made into the abdominal cavity of Bama pigs. An artificial pneumoperitoneum was established using a laparoscopic pneumoperitoneum machine. A bullet model was inserted into the abdominal cavity to build the bullet wound model. After the bullet model was removed, a shrapnel model was inserted into the mid-abdomen to build the shrapnel wound model. The two types of endoscopic system were used to detect, remove bullet model or shrapnel model of the three Bama pigs respectively. The procedure order of the two systems was assigned according to the random number table method. The surgical success, operation time, endoscopy pipeline patency, endoscopic operation satisfaction, adverse events and equipment defects were recorded.

Results:

Three surgeries were performed using the new portable endoscopic system and three other surgeries using the conventional endoscopic system, all of which were successful. The time of the new portable endoscopic system to find and remove the bullet model, and the shrapnel model were 232.33±11.68 s, 300.33±57.70 s, 170.00±44.44 s and 52.67±2.52 s, respectively. The corresponding time of the conventional endoscopic system were 232.67±21.20 s ( t=-0.054, P=0.962), 256.67±67.00 s ( t=0.880, P=0.472), 176.00±52.42 s ( t=-0.111, P=0.922), 58.67±14.84 s ( t=-0.832, P=0.493), respectively. There was no significant difference between the two systems ( P>0.05). The endoscopy tubes of the two endoscopic systems were both smooth. The operator was satisfied with the endoscopic procedures of both endoscopic systems, and no adverse event or device defect occurred.

Conclusion:

The portable endoscopic system proves to be safe and feasible for the diagnosis and treatment of abdominal trauma in animal models.
Key words
Полный текст: 1 База данных: WPRIM Язык: Zh Журнал: Chinese Journal of Digestive Endoscopy Год: 2024 Тип: Article
Полный текст: 1 База данных: WPRIM Язык: Zh Журнал: Chinese Journal of Digestive Endoscopy Год: 2024 Тип: Article