Analysis on HLA B27 results detected by real-time fluorescence PCR method and flow cytometry / 国际检验医学杂志
Article в Zh | WPRIM | ID: wpr-512692
Ответственная библиотека: WPRO
ABSTRACT
Objective To compare the clinical application value of the real-time fluorescence PCR method and flow cytometry method for detecting human leukocyte antigen B27(HLA-B27).Methods Blood HLA-B27 level in 225 patients with suspected ankylosing spondylitis was detected by using real-time fluorescence PCR method and flow cytometry method.The detection results were compared and analyzed between the two methods.Results The results of 95.1% sample were identical detected by two methods without statistically significant difference(P>0.05).Taken the results of flow cytometry as reference, the sensitivity of real-time fluorescence PCR for detecting HLA-B27 was 94%, the specificity was 96%.Gene sequencing was performed if results of a sample detected by two methods were different, which was identical with the result detected by real-time fluorescence PCR.Conclusion Both methods for detecting HLA-B27 all have high sensitivity and specificity.Real-time fluorescence PCR method is more superior to the flow cytometry method in the results accuracy.
Key words