Comparison of the Consistency and Difference of Three Detection Methods in Blood Concentration Monitoring of Mycophenolic Acid / 中国药房
China Pharmacy ; (12): 2885-2889, 2021.
Article в Zh | WPRIM | ID: wpr-906656
Ответственная библиотека: WPRO
ABSTRACT
OBJECTIVE:To investigate the consistency and difference of f luorescence immunochromatographic and liquid chromatography-tandem mass spectrometry (LC-MS/MS)and enzyme multiplied immunoassay technique (EMIT)in the blood concentration monitoring of mycophenolic acid. METHODS :Fluorescence immunochromatography ,LC-MS/MS and EMIT were used to detect the blood concentration of mycophenolic acid in 61 blood samples of children treated with mycophenolate mofetil ester orally at different time points. Kolmogorov-Smirnov method ,Wilcoxon pairing test ,Passing-Bablok regression ,Cusum method,Spearman correlation analysis ,Bland-Altman scatter diagram were adopted for statistical analysis. RESULTS :Blood concentrations of mycophenolic acid ,which were determined by fluorescence immunochromatography ,LC-MS/MS and EMIT , showed non-normal distribution. Passing-Bablok regression analysis showed that regression equation of fluorescence immunochromatography and LC-MS/MS ,fluorescence immunochromatographic method and EMIT were CFI=0.928 3CLC-MS/MS+0.961 7 and CFI=0.880 7CEMIT-0.488 2(FI means fluorescence immunochromatographic ). Spearman correlation analysis showed that the correlation coefficients between fluorescence immunochromatography and LC-MS/MS ,fluorescence immunochromatography and EMIT were 0.968 and 0.929, respectively (P<0.000 1). Bland Altman scatter plot analysis showed that 3.28% of the 358341451@qq.com difference between fluorescence immunochromatography and LC-MS/MS was outside the consistency limit (±1.96SD), and 1.64% of the difference between fluorescence immuno- chromatography and EMIT was outside the consistency limit (± 1.96SD). Wilcoxon pairing test showed that the results of fluorescence immunochromatography were higher than those of LC-MS/MS (Z=3.76,P=0.000 2)and lower than those of EMIT (Z=-5.96,P<0.000 1). CONCLUSIONS :Fluorescence immunochromatography shows good consistency and correlation with LC-MS/MS and EMIT ;the blood concentrations of mycophenolic acid detected by fluorescence immunochromatography were higher than those by LC-MS/MS and lower than those by EMIT . It can be used for bedside rapid detection. When using the test results of different methods for clinical medication ,the differences of test methods need to be considered.
Key words
Полный текст: 1 База данных: WPRIM Тип исследования: Diagnostic_studies Язык: Zh Журнал: China Pharmacy Год: 2021 Тип: Article
Полный текст: 1 База данных: WPRIM Тип исследования: Diagnostic_studies Язык: Zh Журнал: China Pharmacy Год: 2021 Тип: Article