A prospective multicenter randomized controlled clinical trial study of a domestic single-use digital flexible ureteroscope versus a reusable digital flexible ureteroscope for the treatment of upper urinary tract stones / 中华泌尿外科杂志
Chinese Journal of Urology
; (12): 374-378, 2022.
Article
в Zh
| WPRIM
| ID: wpr-933235
Ответственная библиотека:
WPRO
ABSTRACT
Objective:To compare the performance and surgical outcomes of domestic single-use digital flexible ureteroscopes with reusable digital flexible ureteroscopes in treatment of upper urinary stones.Methods:A prospective, single-blind, multicenter and randomized controlled study was performed from September 2018 to June 2019. Eligible patients were randomly assigned, in a ratio of 1∶1, to either experimental group or control group. The inclusion criteria for the study were: aged 18-75 years, solitary upper urinary stone with stone size between 0.8 and 2.0 cm and CT value less than 1 400 HU, negative preoperative urine culture and normal renal function. Exclusion criteria included: patients with acute urinary tract infection, intransitable urethral strictures, impassable ureteropelvic junction obstructions, systemic hemorrhagic disease, coagulation function abnormalities or bleeding tendency, severe hypertension or cardiopulmonary insufficiency, severe hip malformation and difficulty in meeting the demand of operation position and pregnant and lactation women. The device used in the experimental group was a domestic single-use digital flexible ureteroscope, and the device used in the control group was an imported Olympus digital flexible ureteroscope. The qualified rate of clinical comprehensive evaluation (including image quality and operational performance), the rate of device failure, the stone-free rate and the occurrence rate of adverse events (including increase in urine red blood cell and white blood cell counts, postoperative hematuria, nausea, vomiting, dizziness, and fever) in the two groups were recorded.Results:A total of 186 eligible study cases were collected from the People's Hospital of Wuhan University, the First Affiliated Hospital of Xiamen University, and the First Affiliated Hospital of Guangzhou Medical University. 90 cases in the final experimental group and 88 cases in the control group completed the trial and were included in the evaluation. There were no statistically significant differences among age [(48.40±11.36) vs. (47.40±12.53)years old, P=0.594], male to female ratio (62/28 vs. 56/32, P =0.874), BMI [(24.8±2.1) kg/m 2 vs. (25.1±2.0)kg/m 2,P =0.331], hydronephrosis (no/slight vs. mild/severe) (62/28 vs. 65/23, P =0.874), stone location and stone size [(12.8±4.7) mm vs. (11.9±5.2) mm, P =0.227]. There were no significant differences in terms of qualified rate of clinical comprehensive evaluation [98.9% (89/90) vs. 100.0% (88/88), P =0.991], lithotripsy success rate [84.4% (76/90) vs. 84.1% (74/88), P =0.888], device failure/defect rate (both 0%), and the incidence of adverse events [50.0% (45/90) vs. 52.0% (51/88), P =0.894]. The highest incidence of adverse events in two groups was the increase of red blood cells and white blood cells of routine urine after operation. There was no serious adverse event in the experimental group and 1 serious adverse event in the control group. Conclusions:There was no significant difference in image quality, device failure/defect rate, lithotripsy success rate, and adverse event rate between single-use digital flexible ureteroscopes and reusable digital flexible ureteroscopes for lithotripsy of upper ureteral and pelvic stones. Domestic single-use digital flexible ureteroscopes have good safety and effectiveness in the treatment and microscopy of upper urinary tract stones.
Полный текст:
1
База данных:
WPRIM
Тип исследования:
Clinical_trials
Язык:
Zh
Журнал:
Chinese Journal of Urology
Год:
2022
Тип:
Article