We have done a research on mathematical morphology filter in order to eliminate baseline drift in the wave of electrocardiograph (ECG). Marago morphology filter type was chosen and flat structuring element filtering was found to be the best morphology filter modality for eliminating baseline drift, and the ratio of "signal numerical frequency to length of structuring element" determined the attenuation magnitude of the signal eliminated. We have come to the conclusion that the length of flat structuring element ought to be greater than or equal to the width of the signal component to be eliminated. The arithmetical algorithm was simulated with Matlab software, and was transplanted to DSP hardware platform. The result of experiment has shown that the arithmetic operation is simple and the method spends a short time (less than 1 ms) in eliminating the baseline drift of ECG effectively and instantly.
Subject(s)
全文: 1 索引: WPRIM 主要主题: Algorithms / Signal Processing, Computer-Assisted / Artifacts / Electrocardiography / Equipment Design / Methods 限制: Humans 语言: Zh 期刊: Journal of Biomedical Engineering 年: 2010 类型: Article