Agreement between Scheimpflug Camera and the Swept-source Optical Coherence Tomography Measurements in Keratometry and Higher-order Aberrations
Article 在 En | WPRIM | ID: wpr-894650
Responsible library: WPRO
ABSTRACT
Purpose@#To evaluate the compatibility of corneal curvature and astigmatism, and higher-order aberrations (HOAs) measured by the Scheimpflug camera Pentacam HR and the swept-source optical coherence tomography ANTERION. @*Methods@#This prospective study included normal subjects with no ophthalmic history. Steep keratometry (K), flat K, astigmatism and its axis of the anterior and posterior surfaces, total corneal power, and HOAs using the two instruments were compared. To compare the mean values of the measurements, a paired t-test was used. Bland-Altman analysis was applied to assess the agreement between the two devices. @*Results@#Fifty-three eyes of 53 subjects were evaluated. There were statistically significant differences for steep K, astigmatism, and vector J0, J45 in the anterior surface and total corneal power between the two devices (p 0.81, p 0.78, p 0.37, p < 0.05). Most HOAs, except for the horizontal trefoil, showed clinically acceptable agreements. The total root mean square of HOAs was not significantly different between the two devices (p = 0.122). @*Conclusions@#Most of the keratometric values cannot be used interchangeably. However, the vectoral component of astigmatism showed clinically good agreement. Several HOAs have statistically significant differences; however, almost all HOAs showed acceptable agreements, except for the horizontal trefoil.
全文: 1 索引: WPRIM 研究类型: Observational_studies 语言: En 期刊: Korean Journal of Ophthalmology 年: 2021 类型: Article
全文: 1 索引: WPRIM 研究类型: Observational_studies 语言: En 期刊: Korean Journal of Ophthalmology 年: 2021 类型: Article