Laboratory methods for monitoring argatroban in heparin-induced thrombocytopenia.
Guy, Susan; Kitchen, Steve; Hopkins, Barbara; Chunara, Zunaid; Stephenson-Brown, Alex; van Veen, Joost J.
  • Guy S; Sheffield Haemophilia and Thrombosis Centre, Royal Hallamshire Hospital, Sheffield, UK.
  • Kitchen S; Sheffield Haemophilia and Thrombosis Centre, Royal Hallamshire Hospital, Sheffield, UK.
  • Hopkins B; Special Haematology, Leicester Royal Infirmary, Leicester, UK.
  • Chunara Z; Special Haematology, Leicester Royal Infirmary, Leicester, UK.
  • Stephenson-Brown A; Stago UK Ltd, Theale, UK.
  • van Veen JJ; Sheffield Haemophilia and Thrombosis Centre, Royal Hallamshire Hospital, Sheffield, UK.
Int J Lab Hematol ; 44(2): 399-406, 2022 Apr.
Article in English | MEDLINE | ID: covidwho-1511317
ABSTRACT

INTRODUCTION:

The Summary of Product Characteristics for the direct thrombin inhibitor argatroban states monitoring should be by activated partial thromboplastin time (APTT), with a target range of 1.5-3.0 times the patients' baseline APTT. APTT may be influenced by coagulopathies, lupus anticoagulant and raised FVIII levels. Previous studies have shown sensitivity differences of APTT reagents to argatroban. Some recent publications have favoured the use of anti-IIa methods to determine the plasma drug concentration of argatroban. This study aims to compare the anti-IIa assays Hemoclot thrombin inhibitor assay (HTI) and Ecarin chromogenic assay (ECA) alongside the APTT.

METHODS:

Residual plasma of 25 samples from 8 patients (3 with COVID-19 and HIT n = 18, 5 with HIT n = 7) was tested at two sites site A Sysmex CS5100 by HTI and APTT (Actin FS and SynthASil), and also on Stago STA Compact Max ECA and APTT (CK Prest); and site B Stago STA R Max 2 by ECA and APTT (Cephascreen).

RESULTS:

Mean APTT ratio was 1.96 (Actin FS), 1.84 (SynthASil), 1.59 (CK Prest) and 2.48 (Cephascreen). Mean argatroban concentration by HTI was 0.60 µg/mL and by ECA was 0.65 µg/mL (site A) and 0.70 µg/mL (site B). There was a poor correlation to HTI in APTT ratios (range r2  = .0235-0.4181) with stronger correlations between ECA methods to HTI (r2  = .8998 site A, r2  = .8734 site B).

CONCLUSION:

This study confirms previous publications on the unsuitability of APTT and expands this by being multisited and included APTT reagents that use mechanical clot detection. Both anti-IIa methods are more suitable.
Subject(s)
Keywords

Full text: Available Collection: International databases Database: MEDLINE Main subject: Thrombocytopenia / COVID-19 Type of study: Diagnostic study Limits: Humans Language: English Journal: Int J Lab Hematol Journal subject: Hematology Year: 2022 Document Type: Article Affiliation country: Ijlh.13757

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Collection: International databases Database: MEDLINE Main subject: Thrombocytopenia / COVID-19 Type of study: Diagnostic study Limits: Humans Language: English Journal: Int J Lab Hematol Journal subject: Hematology Year: 2022 Document Type: Article Affiliation country: Ijlh.13757