Your browser doesn't support javascript.
AuNP Coupled Rapid Flow-Through Dot-Blot Immuno-Assay for Enhanced Detection of SARS-CoV-2 Specific Nucleocapsid and Receptor Binding Domain IgG.
Sil, Bijon Kumar; Jamiruddin, Mohd Raeed; Haq, Md Ahsanul; Khondoker, Mohib Ullah; Jahan, Nowshin; Khandker, Shahad Saif; Ali, Tamanna; Oishee, Mumtarin Jannat; Kaitsuka, Taku; Mie, Masayasu; Tomizawa, Kazuhito; Kobatake, Eiry; Haque, Mainul; Adnan, Nihad.
  • Sil BK; Gonoshasthaya-RNA Molecular Diagnostic and Research Center, Dhaka, 1205, Bangladesh.
  • Jamiruddin MR; Department of Pharmacy, BRAC University, Dhaka, 1212, Bangladesh.
  • Haq MA; Gonoshasthaya-RNA Molecular Diagnostic and Research Center, Dhaka, 1205, Bangladesh.
  • Khondoker MU; Gonoshasthaya Samaj Vittik Medical College, Savar, Dhaka, 1344, Bangladesh.
  • Jahan N; Gonoshasthaya-RNA Molecular Diagnostic and Research Center, Dhaka, 1205, Bangladesh.
  • Khandker SS; Gonoshasthaya-RNA Molecular Diagnostic and Research Center, Dhaka, 1205, Bangladesh.
  • Ali T; Gonoshasthaya-RNA Molecular Diagnostic and Research Center, Dhaka, 1205, Bangladesh.
  • Oishee MJ; Gonoshasthaya-RNA Molecular Diagnostic and Research Center, Dhaka, 1205, Bangladesh.
  • Kaitsuka T; School of Pharmacy, International University of Health and Welfare, Okawa, Fukuoka, 831-8501, Japan.
  • Mie M; School of Life Science and Technology, Tokyo Institute of Technology, Yokohama, Kanagawa, 226-8502, Japan.
  • Tomizawa K; Department of Molecular Physiology, Graduate School of Medical Sciences, Kumamoto University, Kumamoto, 860-0811, Japan.
  • Kobatake E; School of Life Science and Technology, Tokyo Institute of Technology, Yokohama, Kanagawa, 226-8502, Japan.
  • Haque M; The Unit of Pharmacology, Faculty of Medicine and Defence Health Universiti Pertahanan, Nasional Malaysia (National Defence University of Malaysia), Kuala Lumpur, 57000, Malaysia.
  • Adnan N; Department of Microbiology, Jahangirnagar University, Savar, Dhaka, 1342, Bangladesh.
Int J Nanomedicine ; 16: 4739-4753, 2021.
Artigo em Inglês | MEDLINE | ID: covidwho-1315916
ABSTRACT

BACKGROUND:

Serological tests detecting severe acute respiratory syndrome coronavirus-2 (SARS-CoV-2) are widely used in seroprevalence studies and evaluating the efficacy of the vaccination program. Some of the widely used serological testing techniques are enzyme-linked immune-sorbent assay (ELISA), chemiluminescence immunoassay (CLIA), and lateral flow immunoassay (LFIA). However, these tests are plagued with low sensitivity or specificity, time-consuming, labor-intensive, and expensive. We developed a serological test implementing flow-through dot-blot assay (FT-DBA) for SARS-CoV-2 specific IgG detection, which provides enhanced sensitivity and specificity while being quick to perform and easy to use.

METHODS:

SARS-CoV-2 antigens were immobilized on nitrocellulose membrane to capture human IgG, which was then detected with anti-human IgG conjugated gold nanoparticle (hIgG-AuNP). A total of 181 samples were analyzed in-house. Within which 35 were further evaluated in US FDA-approved CLIA Elecsys SARS-CoV-2 assay. The positive panel consisted of RT-qPCR positive samples from patients with both <14 days and >14 days from the onset of clinical symptoms. The negative panel contained samples collected from the pre-pandemic era dengue patients and healthy donors during the pandemic. Moreover, the sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) of FT-DBA were evaluated against RT-qPCR positive sera. However, the overall efficacies were assessed with sera that seroconverted against either nucleocapsid (NCP) or receptor-binding domain (RBD).

RESULTS:

In-house ELISA selected a total of 81 true seropositive and 100 seronegative samples. The sensitivity of samples with <14 days using FT-DBA was 94.7%, increasing to 100% for samples >14 days. The overall detection sensitivity and specificity were 98.8% and 98%, respectively, whereas the overall PPV and NPV were 99.6% and 99%. Moreover, comparative analysis between in-house ELISA assays and FT-DBA revealed clinical agreement of Cohen's Kappa value of 0.944. The FT-DBA showed sensitivity and specificity of 100% when compared with commercial CLIA kits.

CONCLUSION:

The assay can confirm past SARS-CoV-2 infection with high accuracy within 2 minutes compared to commercial CLIA or in-house ELISA. It can help track SARS-CoV-2 disease progression, population screening, and vaccination response. The ease of use of the assay without requiring any instruments while being semi-quantitative provides the avenue of its implementation in remote areas around the globe, where conventional serodiagnosis is not feasible.
Assuntos
Palavras-chave

Texto completo: Disponível Coleções: Bases de dados internacionais Base de dados: MEDLINE Assunto principal: Imunoglobulina G / Immunoblotting / Nucleocapsídeo / Nanopartículas Metálicas / SARS-CoV-2 / Ouro Tipo de estudo: Estudo diagnóstico / Estudo experimental / Estudo observacional / Estudo prognóstico Tópicos: Vacinas Limite: Adulto / Humanos / Masculino Idioma: Inglês Revista: Int J Nanomedicine Ano de publicação: 2021 Tipo de documento: Artigo País de afiliação: IJN.S313140

Similares

MEDLINE

...
LILACS

LIS


Texto completo: Disponível Coleções: Bases de dados internacionais Base de dados: MEDLINE Assunto principal: Imunoglobulina G / Immunoblotting / Nucleocapsídeo / Nanopartículas Metálicas / SARS-CoV-2 / Ouro Tipo de estudo: Estudo diagnóstico / Estudo experimental / Estudo observacional / Estudo prognóstico Tópicos: Vacinas Limite: Adulto / Humanos / Masculino Idioma: Inglês Revista: Int J Nanomedicine Ano de publicação: 2021 Tipo de documento: Artigo País de afiliação: IJN.S313140