Your browser doesn't support javascript.
loading
Validation of the NucliSens Extractor combined with the AmpliScreen HIV version 1.5 and HCV version 2.0 test for application in NAT minipool screening.
Jongerius, John M; Sjerps, Margret; Cuijpers, H Theo M; van Drimmelen, Harry A J; van der Poel, Cees L; Reesink, Henk W; Molijn, M Hans J; Peeters, Geert A H; Peeters, Ton P W; Lelie, P Nico.
Affiliation
  • Jongerius JM; Sanquin Blood Bank Nortwest, Utrecht, the Netherlands. jmjongerius@hotmail.com
Transfusion ; 42(6): 792-7, 2002 Jun.
Article in En | MEDLINE | ID: mdl-12147034
ABSTRACT

BACKGROUND:

Routine HCV NAT minipool screening (48 donations) of all blood donations was implemented in July 1999 and was combined with HIV NAT in November 2000. This report describes the validation of the NAT methods and the results of quality control testing. STUDY DESIGN AND

METHODS:

Nucleic acid was extracted from 2-mL plasma samples by using an automated silica-based extraction method (NucliSens Extractor, Organon Teknika). Eluates were tested with RT-PCR (AmpliScreen HIV-1 version 1.5 and AmpliScreen HCV version 2.0 test, Roche Diagnostic Systems). HIV-1 and HCV RNA reference panels and run controls (PeliCheck and PeliSpy, respectively, Sanquin-CLB) and human plasma minipools were used for NAT validation.

RESULTS:

The 95-percent detection limit (and 95% CI) for HIV-1 RNA genotype B, HIV-1 RNA genotype E, and HCV RNA genotype 1 was 32 (19-76), 30 (17-72), and 21 (13-44) genome equivalents (geq) per mL, respectively. During initial validation, 2332 samples for HIV-1 RNA and 2644 samples for HCV RNA were analyzed, with 13 (0.56%) and 12 (0.45%) invalid test results, respectively. Thereafter, over 19,600 samples (minipools and run controls) were analyzed during the first 11 months of routine screening. Invalid test results for HIV-1 RNA and HCV RNA were found in 1.1 and 1.07 percent of the samples tested, respectively. HIV-1 RNA minipool testing resulted in 27 (0.16%) initial false-positive results and 3 (0.02%) confirmed positive results. HCV RNA minipool testing resulted in four (0.02%) initial false-positive results and five (0.02%) confirmed positive results.

CONCLUSION:

Routine HIV and HCV NAT minipool screening using the NucliSens Extractor, AmpliScreen HIV-1 version 1.5, and AmpliScreen HCV version 2.0 meets the sensitivity criteria set by the regulatory bodies and provides sufficient specificity and robustness for timely release of blood donations.
Subject(s)
Search on Google
Database: MEDLINE Main subject: Viremia / Blood Donors / RNA, Viral / HIV Infections / Mass Screening / Hepatitis C / Nucleic Acid Amplification Techniques Type of study: Diagnostic_studies / Prognostic_studies / Screening_studies Limits: Humans Country/Region as subject: Europa Language: En Journal: Transfusion Year: 2002 Type: Article
Search on Google
Database: MEDLINE Main subject: Viremia / Blood Donors / RNA, Viral / HIV Infections / Mass Screening / Hepatitis C / Nucleic Acid Amplification Techniques Type of study: Diagnostic_studies / Prognostic_studies / Screening_studies Limits: Humans Country/Region as subject: Europa Language: En Journal: Transfusion Year: 2002 Type: Article