Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 10 de 10
Filter
1.
J Forensic Sci ; 53(3): 632-51, 2008 May.
Article in English | MEDLINE | ID: mdl-18471209

ABSTRACT

An automated process has been developed for the analysis of forensic casework samples using TECAN Genesis RSP 150/8 or Freedom EVO liquid handling workstations equipped exclusively with nondisposable tips. Robot tip cleaning routines have been incorporated strategically within the DNA extraction process as well as at the end of each session. Alternative options were examined for cleaning the tips and different strategies were employed to verify cross-contamination. A 2% sodium hypochlorite wash (1/5th dilution of the 10.8% commercial bleach stock) proved to be the best overall approach for preventing cross-contamination of samples processed using our automated protocol. The bleach wash steps do not adversely impact the short tandem repeat (STR) profiles developed from DNA extracted robotically and allow for major cost savings through the implementation of fixed tips. We have demonstrated that robotic workstations equipped with fixed pipette tips can be used with confidence with properly designed tip washing routines to process casework samples using an adapted magnetic bead extraction protocol.


Subject(s)
Automation , Equipment Contamination/prevention & control , Forensic Medicine , Robotics , Specimen Handling/methods , DNA/analysis , DNA Fingerprinting/methods , Disinfectants , Humans , Laboratories/organization & administration , Magnetics , Microspheres , Polymerase Chain Reaction/methods , Sodium Hypochlorite
2.
Biotechniques ; 59(1): 19-25, 2015 Jul.
Article in English | MEDLINE | ID: mdl-26156780

ABSTRACT

DNA damage in the form of abasic sites, chemically altered nucleotides, and strand fragmentation is the foremost limitation in obtaining genetic information from many ancient samples. Upon cell death, DNA continues to endure various chemical attacks such as hydrolysis and oxidation, but repair pathways found in vivo no longer operate. By incubating degraded DNA with specific enzyme combinations adopted from these pathways, it is possible to reverse some of the post-mortem nucleic acid damage prior to downstream analyses such as library preparation, targeted enrichment, and high-throughput sequencing. Here, we evaluate the performance of two available repair protocols on previously characterized DNA extracts from four mammoths. Both methods use endonucleases and glycosylases along with a DNA polymerase-ligase combination. PreCR Repair Mix increases the number of molecules converted to sequencing libraries, leading to an increase in endogenous content and a decrease in cytosine-to-thymine transitions due to cytosine deamination. However, the effects of Nelson Repair Mix on repair of DNA damage remain inconclusive.


Subject(s)
Bone and Bones/chemistry , DNA Damage , DNA Repair Enzymes/chemistry , DNA/chemistry , High-Throughput Nucleotide Sequencing/methods , Mammoths/genetics , Animals , Fossils
3.
J Forensic Sci ; 49(5): 939-53, 2004 Sep.
Article in English | MEDLINE | ID: mdl-15461094

ABSTRACT

A bioinformatic tool was developed to assist with the victim identification initiative that followed the Swissair Flight 111 disaster. Making use of short tandem repeat (STR) DNA typing data generated with AmpFlSTR Profiler Plus (PP) and AmpFlSTR COfiler(CO) kits, the software systematically compared each available STR genotype with every other genotype. The matching algorithm was based on the search for: (i) direct matches to genotypes derived from personal effects; and (ii) potential kinship associations between victims and next-of-kin, as measured by allele sharing at individual loci. The software greatly assisted parentage analysis by enabling kinship evaluation in situations where complete parentage trios were unavailable and, in some situations, with distantly related relatives. Exclusion of fortuitous kinship associations (FKA) was made possible through the recovery at the disaster site of at least one remains for every sought-after victim, and was incorporated into the kinship software. The data from the 13 combined STR loci produced 6 and 23 times fewer FKAs when compared with PP alone and AmpFlSTR Profiler (PR) alone, respectively. Identification leads or confirmations of identification were obtained for 218 victims for which DNA reference samples (personal effects and kin) had been submitted. Confirmation of an inferred kinship association was sought through frequency and likelihood calculations, as well as corroborative data from other identification modalities. The use of a simple, yet powerful, automated genotype comparison approach and the use of megaplexes with high power of discrimination (PD) values extended considerably the identification capabilities in the case of the Swissair disaster. The DNA typing identification modality proved to be a valuable component of the large arsenal of identification tools deployed in the aftermath of this disaster.


Subject(s)
Accidents, Aviation , DNA Fingerprinting/methods , Tandem Repeat Sequences , Algorithms , Alleles , Female , Genetic Variation , Genotype , Humans , Likelihood Functions , Male , Mutation , Pedigree , Polymerase Chain Reaction , Sex Determination Processes , Software
4.
J Forensic Sci ; 49(5): 968-80, 2004 Sep.
Article in English | MEDLINE | ID: mdl-15461097

ABSTRACT

To assist the interpretation of STR DNA typing results from forensic casework samples containing mixtures, the range of heterozygous allele peak height and peak area ratios (HR) and stutter percentages (stutter %) for the loci comprised in the AmpFlSTR Profiler Plus (PP) kit were assessed on 468 database and 275 casework single source samples. Stutter % medians were similar for database and casework samples, ranging from 2% to 7%. The upper limit of the stutter value range was 16%, calculated as median +3 SD, although lower locus-specific values could be used. HR medians were 93 +/- 6.5% for database samples, 88 +/- 12% for casework samples. For casework samples, the maximum signal imbalance noted was 52%, calculated as median -3 SD. No significant difference was observed between peak height and peak area calculated values. This study shows the importance of selecting the proper reference database for the establishment of HR threshold values.


Subject(s)
Alleles , DNA Fingerprinting/methods , Heterozygote , Tandem Repeat Sequences , Databases as Topic , Genetics, Population , Humans , Polymerase Chain Reaction , Sequence Analysis, DNA
5.
J Forensic Sci ; 48(5): 1014-34, 2003 Sep.
Article in English | MEDLINE | ID: mdl-14535664

ABSTRACT

As part of the validation of the AmpFlSTR Profiler Plus short tandem repeat (STR) system, under reduced polymerase chain reaction (PCR) volume conditions (i.e., 25 microL), a total of 275 casework samples were processed. Examples of profiles are presented along with amplification conditions to improve the odds of obtaining balanced and complete profiles for samples showing partial results or profiles with a descending slope. Data collected and used to develop our interpretation guidelines are included. From the mixture studies, full profiles were obtained for minor contributors, using 2 ng of DNA, with ratios of 10:1 or 1:20 and using 1 ng of DNA, with ratios of 10:1 and 1:8. The specificity of the Profiler Plus amplification reaction performed in 25 microL was examined and confirmed using a large spectrum of nonhuman DNAs. This report supports the use of the AmpFlSTR Profiler Plus STR system for casework DNA typing under reduced PCR volume conditions.


Subject(s)
DNA Fingerprinting/methods , DNA/analysis , Polymerase Chain Reaction/methods , Tandem Repeat Sequences , Alleles , Animals , Bacteria/genetics , Case-Control Studies , Electrophoresis , Female , Forensic Medicine/methods , Humans , Male , Sensitivity and Specificity , Yeasts/genetics
6.
J Forensic Sci ; 48(5): 1001-13, 2003 Sep.
Article in English | MEDLINE | ID: mdl-14535663

ABSTRACT

Improvements in detection limits/sensitivity and lower sample consumption are potential benefits of reducing PCR reaction volumes used in forensic DNA typing of crime scene samples. This premise was studied first with experimental mixtures and a nine-loci megaplex, which demonstrated stochiometric amplification and accurate detection. Next, adjudicated casework samples were subjected to amplification under 15 different template DNA to PCR reaction volume ratios. Reduction of PCR reaction volume and DNA down to 10 microL and 0.500 ng, respectively, produced identical profiles with the same signal intensity and heterozygous allele peak height ratio (HR). Reduction to 5 microL and 0.063 ng yielded HR values that were slightly affected in one to three STR loci. PCR reaction volume reduction can enhance detection and sensitivity while reducing the consumption of irreplaceable crime scene samples.


Subject(s)
DNA Fingerprinting/methods , DNA/analysis , Polymerase Chain Reaction/methods , Tandem Repeat Sequences , Case-Control Studies , Electrophoresis , Feasibility Studies , Female , Forensic Medicine/methods , Humans , Male , Sensitivity and Specificity
7.
Forensic Sci Int Genet ; 6(1): 31-40, 2012 Jan.
Article in English | MEDLINE | ID: mdl-21324769

ABSTRACT

The ability to properly collect, analyze and preserve biological stains is important to preserving the integrity of forensic evidence. Stabilization of intact biological evidence in cells and the DNA extracts from them is particularly important since testing is generally not performed immediately following collection. Furthermore, retesting of stored DNA samples may be needed in casework for replicate testing, confirmation of results, and to accommodate future testing with new technologies. A novel room temperature DNA storage medium, SampleMatrix™ (SM; Biomatrica, Inc., San Diego, CA), was evaluated for stabilizing and protecting samples. Human genomic DNA samples at varying amounts (0.0625-200 ng) were stored dry in SM for 1 day to 1 year under varying conditions that included a typical ambient laboratory environment and also through successive freeze-thaw cycles (3 cycles). In addition, spiking of 1-4 × SM into samples prior to analysis was performed to determine any inhibitory effects of SM. Quantification of recovered DNA following storage was determined by quantitative PCR or by agarose gel electrophoresis, and evaluation of quantitative peak height results from multiplex short tandem repeat (STR) analyses were performed to assess the efficacy of SM for preserving DNA. Results indicate no substantial differences between the quality of samples stored frozen in liquid and those samples maintained dry at ambient temperatures protected in SM. For long-term storage and the storage of low concentration samples, SM provided a significant advantage over freezer storage through higher DNA recovery. No detectable inhibition of amplification was observed at the recommended SM concentration and complete profiles were obtained from genomic DNA samples even in the presence of higher than recommended concentrations of the SM storage medium. The ability to stabilize and protect DNA from degradation at ambient temperatures for extended time periods could have tremendous impact in simplifying and improving sample storage conditions and requirements. The current work focuses on forensics analysis; however this technology is applicable to all endeavors requiring storage of DNA.


Subject(s)
DNA/chemistry , Forensic Genetics , Specimen Handling , Temperature , Humans
8.
Forensic Sci Int Genet ; 4(5): 292-304, 2010 Oct.
Article in English | MEDLINE | ID: mdl-20457033

ABSTRACT

A semi-automated DNA extraction process for casework samples based on the Promega DNA IQ™ system was optimized and validated on TECAN Genesis 150/8 and Freedom EVO robotic liquid handling stations configured with fixed tips and a TECAN TE-Shake™ unit. The use of an orbital shaker during the extraction process promoted efficiency with respect to DNA capture, magnetic bead/DNA complex washes and DNA elution. Validation studies determined the reliability and limitations of this shaker-based process. Reproducibility with regards to DNA yields for the tested robotic workstations proved to be excellent and not significantly different than that offered by the manual phenol/chloroform extraction. DNA extraction of animal:human blood mixtures contaminated with soil demonstrated that a human profile was detectable even in the presence of abundant animal blood. For exhibits containing small amounts of biological material, concordance studies confirmed that DNA yields for this shaker-based extraction process are equivalent or greater to those observed with phenol/chloroform extraction as well as our original validated automated magnetic bead percolation-based extraction process. Our data further supports the increasing use of robotics for the processing of casework samples.


Subject(s)
DNA/isolation & purification , Robotics , Automation , Humans , Reproducibility of Results
9.
Forensic Sci Int Genet ; 4(2): 80-8, 2010 Feb.
Article in English | MEDLINE | ID: mdl-20129465

ABSTRACT

Archival tissue preserved in fixative constitutes an invaluable resource for histological examination, molecular diagnostic procedures and for DNA typing analysis in forensic investigations. However, available material is often limited in size and quantity. Moreover, recovery of DNA is often severely compromised by the presence of covalent DNA-protein cross-links generated by formalin, the most prevalent fixative. We describe the evaluation of buffer formulations, sample lysis regimens and DNA recovery strategies and define optimized manual and automated procedures for the extraction of high quality DNA suitable for molecular diagnostics and genotyping. Using a 3-step enzymatic digestion protocol carried out in the absence of dithiothreitol, we demonstrate that DNA can be efficiently released from cells or tissues preserved in buffered formalin or the alcohol-based fixative GenoFix. This preparatory procedure can then be integrated to traditional phenol/chloroform extraction, a modified manual DNA IQ or automated DNA IQ/Te-Shake-based extraction in order to recover DNA for downstream applications. Quantitative recovery of high quality DNA was best achieved from specimens archived in GenoFix and extracted using magnetic bead capture.


Subject(s)
DNA/genetics , Gene Amplification , Automation/methods , DNA/isolation & purification , DNA, Neoplasm/genetics , Ethanol , Fixatives , Forensic Medicine/methods , Forensic Medicine/trends , Formaldehyde , Hep G2 Cells , Humans , Liver/chemistry , Microsatellite Repeats/genetics , Polymerase Chain Reaction/methods , Tissue Preservation/methods
10.
Electrophoresis ; 25(6): 790-6, 2004 Mar.
Article in English | MEDLINE | ID: mdl-15004837

ABSTRACT

Base-calling precision of short tandem repeat (STR) allelic bands on dynamic slab-gel electrophoresis systems was evaluated. Data was collected from over 6000 population database allele peaks generated from 468 population database samples amplified with the AmpF/STR Profiler Plus (PP) kit and electrophoresed on ABD 377 DNA sequencers. Precision was measured by way of standard deviations and was shown to be essentially the same, whether using fixed or floating bin genotyping. However, the allelic ladders have proven more sensitive to electrophoretic variations than database samples, which have caused some floating bins of D18S51 to shift on occasion. This observation prompted the investigation of polyacrylamide gel formulations in order to stabilize allelic ladder migration. The results demonstrate that, although alleles comprised in allelic ladders and questioned samples run on the same gel should migrate in an identical manner, this premise needs to be verified for any given electrophoresis platform and gel formulation. We show that the compilation of base-calling data is a very informative and useful tool for assessing the performance stability of dynamic gel electrophoresis systems, stability on which depends genotyping result quality.


Subject(s)
Alleles , DNA Fingerprinting/methods , Electrophoresis, Polyacrylamide Gel , Forensic Medicine/methods , Tandem Repeat Sequences/genetics , Fluorescence , Humans , Polymerase Chain Reaction , Reproducibility of Results , Sequence Analysis, DNA
SELECTION OF CITATIONS