J Biomed Opt ; 13(4): 044016, 2008.
Article in English | MEDLINE | ID: mdl-19021344

ABSTRACT

We describe the design and operation of a multispectral confocal microendoscope. This fiber-based fluorescence imaging system consists of a slit-scan confocal microscope coupled to an imaging catheter that is designed to be minimally invasive and allow for cellular level imaging in vivo. The system can operate in two imaging modes. The grayscale mode of operation provides high resolution real-time in vivo images showing the intensity of fluorescent signal from the specimen. The multispectral mode of operation uses a prism as a dispersive element to collect a full multispectral image of the fluorescence emission. The instrument can switch back and forth nearly instantaneously between the two imaging modes (less than half a second). In the current configuration, the multispectral confocal microendoscope achieves 3-microm lateral resolution and 30-microm axial resolution. The system records light from 500 to 750 nm, and the minimum resolvable wavelength difference varies from 2.9 to 8.3 nm over this spectral range. Grayscale and multispectral imaging results from ex-vivo human tissues and small animal tissues are presented.


Subject(s)
Endoscopes , Image Enhancement/instrumentation , Microscopy, Confocal/instrumentation , Optical Fibers , Spectrum Analysis/instrumentation , Animals , Equipment Design , Equipment Failure Analysis , Humans , Miniaturization , Reproducibility of Results , Sensitivity and Specificity