Your browser doesn't support javascript.
loading
Feasibility studies of Bragg probe for noninvasive carotid pulse waveform assessment.
Leitão, Cátia; Bilro, Lúcia; Alberto, Nélia; Antunes, Paulo; Lima, Hugo; André, Paulo S; Nogueira, Rogério; Pinto, João L.
Affiliation
  • Leitão C; University of Aveiro, Campus Universitário de Santiago, Department of Physics and I3N, 3810-193 Aveiro, Portugal. catia.leitao@ua.pt
J Biomed Opt ; 18(1): 17006, 2013 Jan.
Article in En | MEDLINE | ID: mdl-23296087
ABSTRACT
The arterial stiffness evaluation is largely reported as an independent predictor of cardiovascular diseases. The central pulse waveform can provide important data about arterial health and has been studied in patients with several pathologies, such as diabetes mellitus, coronary artery disease and hypertension. The implementation and feasibility studies of a fiber Bragg grating probe for noninvasive monitoring of the carotid pulse are described based on fiber Bragg grating technology. Assessment tests were carried out in carotids of different volunteers and it was possible to detect the carotid pulse waveform in all subjects. In one of the subjects, the sensor was also tested in terms of repeatability. Although further tests will be required for clinical investigation, the first studies suggest that the developed sensor can be a valid alternative to electromechanical tonometers.
Subject(s)

Full text: 1 Database: MEDLINE Main subject: Signal Processing, Computer-Assisted / Carotid Arteries / Fiber Optic Technology / Pulse Wave Analysis Type of study: Health_technology_assessment / Prognostic_studies Limits: Humans Language: En Year: 2013 Type: Article

Full text: 1 Database: MEDLINE Main subject: Signal Processing, Computer-Assisted / Carotid Arteries / Fiber Optic Technology / Pulse Wave Analysis Type of study: Health_technology_assessment / Prognostic_studies Limits: Humans Language: En Year: 2013 Type: Article