Your browser doesn't support javascript.
loading
Specification of absorbed dose to water using model-based dose calculation algorithms for treatment planning in brachytherapy.
Tedgren, Åsa Carlsson; Carlsson, Gudrun Alm.
Affiliation
  • Tedgren ÅC; Radiation Physics, Department of Medical and Health Sciences, Linköping University and Center of Medical Image Science and Visualization, SE-581 85 Linköping, Sweden. asa.carlsson.tedgren@liu.se
Phys Med Biol ; 58(8): 2561-79, 2013 Apr 21.
Article in En | MEDLINE | ID: mdl-23528349
ABSTRACT
Model-based dose calculation algorithms (MBDCAs), recently introduced in treatment planning systems (TPS) for brachytherapy, calculate tissue absorbed doses. In the TPS framework, doses have hereto been reported as dose to water and water may still be preferred as a dose specification medium. Dose to tissue medium Dmed then needs to be converted into dose to water in tissue Dw,med. Methods to calculate absorbed dose to differently sized water compartments/cavities inside tissue, infinitesimal (used for definition of absorbed dose), small, large or intermediate, are reviewed. Burlin theory is applied to estimate photon energies at which cavity sizes in the range 1 nm-10 mm can be considered small or large. Photon and electron energy spectra are calculated at 1 cm distance from the central axis in cylindrical phantoms of bone, muscle and adipose tissue for 20, 50, 300 keV photons and photons from (125)I, (169)Yb and (192)Ir sources; ratios of mass-collision-stopping powers and mass energy absorption coefficients are calculated as applicable to convert Dmed into Dw,med for small and large cavities. Results show that 1-10 nm sized cavities are small at all investigated photon energies; 100 µm cavities are large only at photon energies <20 keV. A choice of an appropriate conversion coefficient Dw, med/Dmed is discussed in terms of the cavity size in relation to the size of important cellular targets. Free radicals from DNA bound water of nanometre dimensions contribute to DNA damage and cell killing and may be the most important water compartment in cells implying use of ratios of mass-collision-stopping powers for converting Dmed into Dw,med.
Subject(s)

Full text: 1 Database: MEDLINE Main subject: Radiation Dosage / Algorithms / Radiotherapy Planning, Computer-Assisted / Brachytherapy / Water / Models, Theoretical Type of study: Health_economic_evaluation Limits: Humans Language: En Year: 2013 Type: Article

Full text: 1 Database: MEDLINE Main subject: Radiation Dosage / Algorithms / Radiotherapy Planning, Computer-Assisted / Brachytherapy / Water / Models, Theoretical Type of study: Health_economic_evaluation Limits: Humans Language: En Year: 2013 Type: Article