Your browser doesn't support javascript.
loading
Commissioning of the world's first compact pencil-beam scanning proton therapy system.
Pidikiti, Rajesh; Patel, Bijal C; Maynard, Matthew R; Dugas, Joseph P; Syh, Joseph; Sahoo, Narayan; Wu, Hsinshun Terry; Rosen, Lane R.
Affiliation
  • Pidikiti R; Procure Proton Therapy Center, Oklahoma City, OK, USA.
  • Patel BC; Radiation Oncology, Methodist Hospital, Baytown, TX, USA.
  • Maynard MR; Radiation Oncology, Willis-Knighton Cancer Center, Shreveport, LA, USA.
  • Dugas JP; Radiation Oncology, Willis-Knighton Cancer Center, Shreveport, LA, USA.
  • Syh J; Radiation Oncology, Willis-Knighton Cancer Center, Shreveport, LA, USA.
  • Sahoo N; Division of Radiation Oncology, Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, TX, USA.
  • Wu HT; Radiation Oncology, Willis-Knighton Cancer Center, Shreveport, LA, USA.
  • Rosen LR; Radiation Oncology, Willis-Knighton Cancer Center, Shreveport, LA, USA.
J Appl Clin Med Phys ; 19(1): 94-105, 2018 Jan.
Article in En | MEDLINE | ID: mdl-29152838
ABSTRACT
This paper summarizes clinical commissioning of the world's first commercial, clinically utilized installation of a compact, image-guided, pencil-beam scanning, intensity-modulated proton therapy system, the IBA Proteus® ONE, at the Willis-Knighton Cancer Center (WKCC) in Shreveport, LA. The Proteus® ONE is a single-room, compact-gantry system employing a cyclotron-generated proton beam with image guidance via cone-beam CT as well as stereoscopic orthogonal and oblique planar kV imaging. Coupling 220° of gantry rotation with a 6D robotic couch capable of in plane patient rotations of over 180° degrees allows for 360° of treatment access. Along with general machine characterization, system commissioning required (a) characterization and calibration of the proton beam, (b) treatment planning system commissioning including CT-to-density curve determination, (c) image guidance system commissioning, and (d) safety verification (interlocks and radiation survey). System readiness for patient treatment was validated by irradiating calibration TLDs as well as prostate, head, and lung phantoms from the Imaging and Radiation Oncology Core (IROC), Houston. These results confirmed safe and accurate machine functionality suitable for patient treatment. WKCC also successfully completed an on-site dosimetry review by an independent team of IROC physicists that corroborated accurate Proteus® ONE dosimetry.
Subject(s)
Key words

Full text: 1 Database: MEDLINE Main subject: Prostatic Neoplasms / Radiotherapy Planning, Computer-Assisted / Phantoms, Imaging / Proton Therapy / Head and Neck Neoplasms / Lung Neoplasms Type of study: Guideline / Qualitative_research Limits: Humans / Male Language: En Year: 2018 Type: Article

Full text: 1 Database: MEDLINE Main subject: Prostatic Neoplasms / Radiotherapy Planning, Computer-Assisted / Phantoms, Imaging / Proton Therapy / Head and Neck Neoplasms / Lung Neoplasms Type of study: Guideline / Qualitative_research Limits: Humans / Male Language: En Year: 2018 Type: Article