Outcomes of Micra leadless pacemaker implantation with uninterrupted anticoagulation.
Kiani, Soroosh; Black, George B; Rao, Birju; Thakkar, Nancy; Massad, Christopher; Patel, Akshar V; Merchant, Faisal M; Hoskins, Michael H; De Lurgio, David B; Patel, Anshul M; Shah, Anand D; Leon, Angel R; Westerman, Stacy B; Lloyd, Michael S; El-Chami, Mikhael F.
Affiliation
  • Kiani S; Division of Cardiovascular Disease, Department of Medicine, Emory University School of Medicine, Atlanta, Georgia.
  • Black GB; Division of Cardiovascular Disease, Department of Medicine, Emory University School of Medicine, Atlanta, Georgia.
  • Rao B; Division of Cardiovascular Disease, Department of Medicine, Emory University School of Medicine, Atlanta, Georgia.
  • Thakkar N; Division of Cardiovascular Disease, Department of Medicine, Emory University School of Medicine, Atlanta, Georgia.
  • Massad C; Division of Cardiovascular Disease, Department of Medicine, Emory University School of Medicine, Atlanta, Georgia.
  • Patel AV; Division of Cardiovascular Disease, Department of Medicine, Emory University School of Medicine, Atlanta, Georgia.
  • Merchant FM; Division of Cardiovascular Disease, Department of Medicine, Emory University School of Medicine, Atlanta, Georgia.
  • Hoskins MH; Division of Cardiovascular Disease, Department of Medicine, Emory University School of Medicine, Atlanta, Georgia.
  • De Lurgio DB; Division of Cardiovascular Disease, Department of Medicine, Emory University School of Medicine, Atlanta, Georgia.
  • Patel AM; Division of Cardiovascular Disease, Department of Medicine, Emory University School of Medicine, Atlanta, Georgia.
  • Shah AD; Division of Cardiovascular Disease, Department of Medicine, Emory University School of Medicine, Atlanta, Georgia.
  • Leon AR; Division of Cardiovascular Disease, Department of Medicine, Emory University School of Medicine, Atlanta, Georgia.
  • Westerman SB; Division of Cardiovascular Disease, Department of Medicine, Emory University School of Medicine, Atlanta, Georgia.
  • Lloyd MS; Division of Cardiovascular Disease, Department of Medicine, Emory University School of Medicine, Atlanta, Georgia.
  • El-Chami MF; Division of Cardiovascular Disease, Department of Medicine, Emory University School of Medicine, Atlanta, Georgia.
J Cardiovasc Electrophysiol ; 30(8): 1313-1318, 2019 08.
Article in En | MEDLINE | ID: mdl-31045296
ABSTRACT

BACKGROUND:

Implantation of the MICRA Leadless pacemaker requires the use of a 27 French introducer, blunt delivery system and device fixation to the myocardium via nitinol tines. While prior studies have proven its safety, it is unclear whether performing this procedure with uninterrupted anticoagulation exposes patients to increased risks. We sought to investigate the feasibility and safety of continuing therapeutic anticoagulation during the periprocedural period.

METHODS:

We evaluated all patients undergoing MICRA placement at our institution between April 2014 and August 2018 with complete follow-up data (n = 170). Patients were stratified into two groups those on active anticoagulation (OAC, n = 26), defined as having an International normalized ratio >2.0 or having continued a direct oral anticoagulant, and those not anticoagulated (Off-OAC, n = 144). We evaluated for a composite outcome of all major complications, including access site complications and pericardial effusion.

RESULTS:

OAC and Off-OAC groups had similar mean age (74 ± 13 vs 75 ± 13 years; P = .914). The OAC group had a nonsignificantly lower prevalence of end-stage renal disease (8% vs 17%; P = .375) and aspirin use (27% vs 47%; P = .131). Those in the OAC group were more likely to be on warfarin than those in the Off-OAC group (81% vs 30%; P < .001). The rate of the composite endpoint was similar between the OAC and Off-OAC groups (3.8 % vs 1.4%, respectively; P = .761). Length of stay was similar between groups (1.3 ± 2.6 vs 2.3 ± 3.4 days; P = 0.108).

CONCLUSION:

Continuation of therapeutic anticoagulation during MICRA implantation appears to be feasible, safe and associated with shorter hospitalization among appropriately selected individuals.
Subject(s)
Key words

Full text: 1 Database: MEDLINE Main subject: Pacemaker, Artificial / Arrhythmias, Cardiac / Cardiac Pacing, Artificial / Heart Failure / Anticoagulants Type of study: Diagnostic_studies / Etiology_studies / Observational_studies / Risk_factors_studies Limits: Aged / Aged80 / Female / Humans / Male / Middle aged Language: En Year: 2019 Type: Article

Full text: 1 Database: MEDLINE Main subject: Pacemaker, Artificial / Arrhythmias, Cardiac / Cardiac Pacing, Artificial / Heart Failure / Anticoagulants Type of study: Diagnostic_studies / Etiology_studies / Observational_studies / Risk_factors_studies Limits: Aged / Aged80 / Female / Humans / Male / Middle aged Language: En Year: 2019 Type: Article