Your browser doesn't support javascript.
loading
Cardiac 123I-MIBG normal uptake values are population-specific: Results from a cohort of controls over 60 years of age.
Roberts, G; Lloyd, J J; Kane, J P M; Durcan, R; Lawley, S; Howe, K; Petrides, G S; O'Brien, J T; Thomas, A J.
Affiliation
  • Roberts G; Institute of Neuroscience, Newcastle University, Biomedical Research Building, Campus for Ageing and Vitality, Westgate Road, Newcastle upon Tyne, NE4 6BE, UK. gemma.roberts@newcastle.ac.uk.
  • Lloyd JJ; Nuclear Medicine Department, Leazes Wing, Royal Victoria Infirmary, Richardson Road, Newcastle upon Tyne, NE1 4LP, UK. gemma.roberts@newcastle.ac.uk.
  • Kane JPM; Institute of Neuroscience, Newcastle University, Biomedical Research Building, Campus for Ageing and Vitality, Westgate Road, Newcastle upon Tyne, NE4 6BE, UK.
  • Durcan R; Nuclear Medicine Department, Leazes Wing, Royal Victoria Infirmary, Richardson Road, Newcastle upon Tyne, NE1 4LP, UK.
  • Lawley S; Institute of Neuroscience, Newcastle University, Biomedical Research Building, Campus for Ageing and Vitality, Westgate Road, Newcastle upon Tyne, NE4 6BE, UK.
  • Howe K; Institute of Neuroscience, Newcastle University, Biomedical Research Building, Campus for Ageing and Vitality, Westgate Road, Newcastle upon Tyne, NE4 6BE, UK.
  • Petrides GS; Institute of Neuroscience, Newcastle University, Biomedical Research Building, Campus for Ageing and Vitality, Westgate Road, Newcastle upon Tyne, NE4 6BE, UK.
  • O'Brien JT; Nuclear Medicine Department, Leazes Wing, Royal Victoria Infirmary, Richardson Road, Newcastle upon Tyne, NE1 4LP, UK.
  • Thomas AJ; Nuclear Medicine Department, Leazes Wing, Royal Victoria Infirmary, Richardson Road, Newcastle upon Tyne, NE1 4LP, UK.
J Nucl Cardiol ; 28(4): 1692-1701, 2021 08.
Article in En | MEDLINE | ID: mdl-31529384
ABSTRACT

PURPOSE:

Cardiac 123I-MIBG image interpretation is affected by population differences and technical factors. We recruited older adults without cognitive decline and compared their cardiac MIBG uptake with results from the literature.

METHODS:

Phantom calibration confirmed that cardiac uptake results from Japan could be applied to our center. We recruited 31 controls, 17 individuals with dementia with Lewy bodies (DLB) and 15 with Alzheimer's disease (AD). Images were acquired 20 minutes and four hours after injection using Siemens cameras with medium-energy low-penetration (MELP) collimators. Local normal heart-to-mediastinum (HMR) ratios were compared to Japanese results.

RESULTS:

Siemens gamma cameras with MELP collimators should give HMRs very close to the calibrated values used in Japan. However, our cut-offs with controls were lower at 2.07 for early and 1.86 for delayed images. Applying our lower cut-off to the dementia patients may increase the specificity of cardiac MIBG imaging for DLB diagnosis in a UK population without reducing sensitivity.

CONCLUSIONS:

Our local HMR cut-off values are lower than in Japan, higher than in a large US study but similar to those found in another UK center. UK centers using other cameras and collimators may need to use different cut-offs to apply our results.
Subject(s)
Key words

Full text: 1 Database: MEDLINE Main subject: Radiopharmaceuticals / 3-Iodobenzylguanidine / Lewy Body Disease / Alzheimer Disease / Iodine Radioisotopes Type of study: Etiology_studies / Incidence_studies / Observational_studies / Risk_factors_studies Limits: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Language: En Year: 2021 Type: Article

Full text: 1 Database: MEDLINE Main subject: Radiopharmaceuticals / 3-Iodobenzylguanidine / Lewy Body Disease / Alzheimer Disease / Iodine Radioisotopes Type of study: Etiology_studies / Incidence_studies / Observational_studies / Risk_factors_studies Limits: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Language: En Year: 2021 Type: Article