Super-resolution radar imaging for breast cancer detection with microwaves: the integrated information selection criteria.
Annu Int Conf IEEE Eng Med Biol Soc ; 2019: 1868-1874, 2019 Jul.
Article in En | MEDLINE | ID: mdl-31946262
ABSTRACT
This paper focuses on the data preprocessing scheme, as well as on the frequency selection and spatial filtering modules integrated with a Time-Reversal Multiple Signal Classification (TR-MUSIC) algorithm, for microwave breast imaging. This algorithm is part of the data processing chain of the Wavelia Microwave Breast Imaging (MBI) system prototype, which has been recently installed at the University Hospital of Galway, Ireland, for a first-in-human clinical trial. Indicative results from application of the algorithm on an experimental phantom dataset, and on a first patient dataset, are presented in this paper. Good correspondence between the two datasets is demonstrated, confirming the validity of the experimental setup used so far for the on-site acceptance of the Wavelia system, after installation at the hospital for clinical testing.
Subject(s)

Full text: 1 Database: MEDLINE Main subject: Radar / Breast / Breast Neoplasms / Microwaves Type of study: Diagnostic_studies Limits: Humans Language: En Year: 2019 Type: Article