Your browser doesn't support javascript.
loading
Relationship Between Electrocochleography, Angular Insertion Depth, and Cochlear Implant Speech Perception Outcomes.
Canfarotta, Michael W; O'Connell, Brendan P; Giardina, Christopher K; Buss, Emily; Brown, Kevin D; Dillon, Margaret T; Rooth, Meredith A; Pillsbury, Harold C; Buchman, Craig A; Adunka, Oliver F; Fitzpatrick, Douglas C.
Affiliation
  • Canfarotta MW; Department of Otolaryngology/Head and Neck Surgery, University of North Carolina at Chapel Hill, North Carolina, USA.
  • O'Connell BP; Department of Otolaryngology/Head and Neck Surgery, University of North Carolina at Chapel Hill, North Carolina, USA.
  • Giardina CK; Department of Otolaryngology/Head and Neck Surgery, University of North Carolina at Chapel Hill, North Carolina, USA.
  • Buss E; Department of Otolaryngology/Head and Neck Surgery, University of North Carolina at Chapel Hill, North Carolina, USA.
  • Brown KD; Department of Otolaryngology/Head and Neck Surgery, University of North Carolina at Chapel Hill, North Carolina, USA.
  • Dillon MT; Department of Otolaryngology/Head and Neck Surgery, University of North Carolina at Chapel Hill, North Carolina, USA.
  • Rooth MA; Department of Otolaryngology/Head and Neck Surgery, University of North Carolina at Chapel Hill, North Carolina, USA.
  • Pillsbury HC; Department of Otolaryngology/Head and Neck Surgery, University of North Carolina at Chapel Hill, North Carolina, USA.
  • Buchman CA; Department of Otolaryngology-Head and Neck Surgery, Washington University School of Medicine, St. Louis, Missouri, USA.
  • Adunka OF; Department of Otolaryngology-Head and Neck Surgery, The Ohio State University, Columbus, Ohio, USA.
  • Fitzpatrick DC; Department of Otolaryngology/Head and Neck Surgery, University of North Carolina at Chapel Hill, North Carolina, USA.
Ear Hear ; 42(4): 941-948, 2021.
Article in En | MEDLINE | ID: mdl-33369942
OBJECTIVES: Electrocochleography (ECochG), obtained before the insertion of a cochlear implant (CI) array, provides a measure of residual cochlear function that accounts for a substantial portion of variability in postoperative speech perception outcomes in adults. It is postulated that subsequent surgical factors represent independent sources of variance in outcomes. Prior work has demonstrated a positive correlation between angular insertion depth (AID) of straight arrays and speech perception under the CI-alone condition, with an inverse relationship observed for precurved arrays. The purpose of the present study was to determine the combined effects of ECochG, AID, and array design on speech perception outcomes. DESIGN: Participants were 50 postlingually deafened adult CI recipients who received one of three straight arrays (MED-EL Flex24, MED-EL Flex28, and MED-EL Standard) and two precurved arrays (Cochlear Contour Advance and Advanced Bionics HiFocus Mid-Scala). Residual cochlear function was determined by the intraoperative ECochG total response (TR) measured before array insertion, which is the sum of magnitudes of spectral components in response to tones of different stimulus frequencies across the speech spectrum. The AID was then determined with postoperative imaging. Multiple linear regression was used to predict consonant-nucleus-consonant (CNC) word recognition in the CI-alone condition at 6 months postactivation based on AID, TR, and array design. RESULTS: Forty-one participants received a straight array and nine received a precurved array. The AID of the most apical electrode contact ranged from 341° to 696°. The TR measured by ECochG accounted for 43% of variance in speech perception outcomes (p < 0.001). A regression model predicting CNC word scores with the TR tended to underestimate the performance for precurved arrays and deeply inserted straight arrays, and to overestimate the performance for straight arrays with shallower insertions. When combined in a multivariate linear regression, the TR, AID, and array design accounted for 72% of variability in speech perception outcomes (p < 0.001). CONCLUSIONS: A model of speech perception outcomes that incorporates TR, AID, and array design represents an improvement over a model based on TR alone. The success of this model shows that peripheral factors including cochlear health and electrode placement may play a predominant role in speech perception with CIs.
Subject(s)

Full text: 1 Database: MEDLINE Main subject: Speech Perception / Cochlear Implants / Cochlear Implantation Type of study: Prognostic_studies Limits: Adult / Humans Language: En Year: 2021 Type: Article

Full text: 1 Database: MEDLINE Main subject: Speech Perception / Cochlear Implants / Cochlear Implantation Type of study: Prognostic_studies Limits: Adult / Humans Language: En Year: 2021 Type: Article