Your browser doesn't support javascript.
loading
Insensitivity of machine log files to MLC leaf backlash and effect of MLC backlash on clinical dynamic MLC motion: An experimental investigation.
Barnes, Michael; Pomare, Dennis; Doebrich, Marcus; Standen, Therese S; Wolf, Joshua; Greer, Peter; Simpson, John.
Affiliation
  • Barnes M; Department of Radiation Oncology, Calvary Mater Hospital Newcastle, Newcastle, New South Wales, Australia.
  • Pomare D; School of Mathematical and Physical Sciences, University of Newcastle, Newcastle, New South Wales, Australia.
  • Doebrich M; Department of Radiation Oncology, Calvary Mater Hospital Newcastle, Newcastle, New South Wales, Australia.
  • Standen TS; Department of Radiation Oncology, Calvary Mater Hospital Newcastle, Newcastle, New South Wales, Australia.
  • Wolf J; Department of Radiation Oncology, Calvary Mater Hospital Newcastle, Newcastle, New South Wales, Australia.
  • Greer P; Department of Radiation Oncology, Calvary Mater Hospital Newcastle, Newcastle, New South Wales, Australia.
  • Simpson J; Icon Cancer Centre Maitland, Maitland Private Hospital, Maitland, New South Wales, Australia.
J Appl Clin Med Phys ; 23(9): e13660, 2022 Sep.
Article in En | MEDLINE | ID: mdl-35678793
PURPOSE: Multi-leaf-collimator (MLC) leaf position accuracy is important for accurate dynamic radiotherapy treatment plan delivery. Machine log files have become widely utilized for quality assurance (QA) of such dynamic treatments. The primary aim is to test the sensitivity of machine log files in comparison to electronic portal imaging device (EPID)-based measurements to MLC position errors caused by leaf backlash. The secondary aim is to investigate the effect of MLC leaf backlash on MLC leaf motion during clinical dynamic plan delivery. METHODS: The sensitivity of machine log files and two EPID-based measurements were assessed via a controlled experiment, whereby the length of the "T" section of a series of 12 MLC leaf T-nuts in a Varian Millennium MLC for a Trilogy C-series type linac was reduced by sandpapering the top of the "T" to introduce backlash. The built-in machine MLC leaf backlash test as well as measurements for two EPID-based dynamic MLC positional tests along with log files were recorded pre- and post-T-nut modification. All methods were investigated for sensitivity to the T-nut change by assessing the effect on measured MLC leaf positions. A reduced version of the experiment was repeated on a TrueBeam type linac with Millennium MLC. RESULTS: No significant differences before and after T-nut modification were detected in any of the log file data. Both EPID methods demonstrated sensitivity to the introduced change at approximately the expected magnitude with a strong dependence observed with gantry angle. EPID-based data showed MLC positional error in agreement with the micrometer measured T-nut length change to 0.07 ± 0.05 mm (1 SD) using the departmental routine QA test. Backlash results were consistent between linac types. CONCLUSION: Machine log files appear insensitive to MLC position errors caused by MLC leaf backlash introduced via the T-nut. The effect of backlash on clinical MLC motions is heavily gantry angle dependent.
Subject(s)
Key words

Full text: 1 Database: MEDLINE Main subject: Radiotherapy, Intensity-Modulated Limits: Humans Language: En Year: 2022 Type: Article

Full text: 1 Database: MEDLINE Main subject: Radiotherapy, Intensity-Modulated Limits: Humans Language: En Year: 2022 Type: Article