Subcutaneous cardioverter defibrillator implanted intermuscularly in patients with end-stage renal disease requiring hemodialysis: 5-year follow-up.
Kloppe, Axel; Winter, Joachim; Prull, Magnus; Aweimer, Assem; El-Battrawy, Ibrahim; Hanefeld, Christoph; O'Connor, Stephen; Mügge, Andreas; Schiedat, Fabian.
Affiliation
  • Kloppe A; Department of Cardiology, Intensive Medicine and Angiology, Marienhospital Gelsenkirchen, Academic Hospital of the Ruhr University Bochum, Virchowstraße 122, 45886, Gelsenkirchen, Germany.
  • Winter J; Department of Cardiology and Angiology, University Hospital Bergmannsheil Bochum, Ruhr University Bochum, Bochum, Germany.
  • Prull M; Department of Cardiovascular Surgery, Heinrich Heine University, Dusseldorf, Germany.
  • Aweimer A; Department of Cardiology, Augusta Hospital Bochum, Bochum, Germany.
  • El-Battrawy I; Department of Cardiology and Angiology, University Hospital Bergmannsheil Bochum, Ruhr University Bochum, Bochum, Germany.
  • Hanefeld C; Department of Cardiology and Angiology, University Hospital Bergmannsheil Bochum, Ruhr University Bochum, Bochum, Germany.
  • O'Connor S; Department of Molecular and Experimental Cardiology, Institut Für Forschung Und Lehre (IFL), Ruhr University Bochum, Bochum, Germany.
  • Mügge A; Department of Cardiology, Katholische Kliniken Bochum, Ruhr University Bochum, Bochum, Germany.
  • Schiedat F; Department of Biomedical Engineering, City, University of London, London, UK.
Article in En | MEDLINE | ID: mdl-38383674
ABSTRACT

BACKGROUND:

The aim of the present study was to evaluate the long-term safety and effectiveness of the subcutaneous implantable cardioverter defibrillator (S-ICD) when implanted intermuscularly in patients with end-stage renal disease and hemodialysis.

METHODS:

This study is a retrospective analysis of 21 consecutive patients implanted with S-ICDs at three experienced centers in Germany with comorbid renal insufficiency requiring hemodialysis, as well as being at risk of sudden cardiac death. The S-ICD was placed intermuscularly in all patients. Follow-ups (FUs) were performed every 6 months.

RESULTS:

The mean ± standard deviation FU duration was 60.0 ± 11.4 months, with a range of 39 to 78 months. There were no deaths due to arrhythmia, or device-associated infections and complications. Four patients (19.1%) died during FU due to respiratory insufficiency during dialysis, systolic heart failure, septic infection of the urogenital tract, and colorectal cancer, respectively. There were six non-device-related hospitalizations with a duration of 12.7 ± 5.1 days and a hospitalization rate of 4.1 per 100 patient years.

CONCLUSIONS:

In the long-term FU of this small population of seriously compromised hemodialysis patients at risk of sudden cardiac death, the intermuscularly implanted S-ICD system was safe and effective. No arrhythmic complications, device-associated infections, or complications compromised survival. These data are encouraging and support testing in a larger group of similarly compromised patients.