Your browser doesn't support javascript.
loading
Comparison of the sensitivity of commercial APTT reagents in the detection of mild coagulopathies.
Am J Clin Pathol ; 82(4): 436-9, 1984 Oct.
Article in En | MEDLINE | ID: mdl-6475840
ABSTRACT
This study was undertaken to evaluate the precision and sensitivity of three different commercial APTT reagents containing the activators kaolin, micronized silica, or ellagic acid. These reagents varied greatly in their ability to detect mild coagulopathies. The ellagic acid reagent was able to detect the mildest deficiencies for the most common coagulopathies. This reagent was sensitive to 50% levels of Factor VIIIC, whereas the APTT with the kaolin reagent was not prolonged until levels of 35% or less were attained. The micronized silica reagent was the least sensitive to Factor IX deficiency, detecting levels of 12% or less. Precision was similar for all reagents when tested with normal and slightly abnormal plasmas. Since methods and instrumentation vary, each laboratory should evaluate their APTT reagent to determine its precision and sensitivity.
Subject(s)
Search on Google
Database: MEDLINE Main subject: Partial Thromboplastin Time / Blood Coagulation Disorders / Blood Coagulation Tests / Indicators and Reagents Type of study: Diagnostic_studies / Evaluation_studies Limits: Female / Humans / Male Language: En Year: 1984 Type: Article