Suitability of the upper airway models obtained from MRI studies in simulating drug lung deposition from inhalers.
Pharm Res
; 22(1): 166-70, 2005 Jan.
Article
en En
| MEDLINE
| ID: mdl-15771244
ABSTRACT
PURPOSE:
In this study, the suitability of the upper airway models, obtained by applying a magnetic resonance imaging method, in simulating in vivo aerosol deposition data is determined.METHODS:
Depositions of salbutamol sulfate from two nebulizers in two models, one with constriction at the oropharynx (the constricted cast) and another model without that constriction (the wide cast), were determined.RESULTS:
For the Sidestream and Ventstream nebulizer, 76 +/- 3% (mean +/- standard deviation) and 81 +/- 2% of the emitted dose deposited in the constricted cast, whereas 51 +/- 2% and 49 +/- 3% of the emitted dose deposited in the wide cast, respectively. These values were in good agreement with in vivo data. Mostly, increasing nebulizer charge volume (by normal saline) from 2.5 ml to 5 ml increased significantly the lung dose. However, the lung doses from the Sidestream and Ventstream nebulizer with 2.5 ml charge volume via the wide cast were (1.37 +/- 0.06 and 1.38 +/- 0.05 mg) significantly larger than those for the constricted cast with 5 ml charge volume (0.87 +/- 0.15 and 0.86 +/- 0.21 mg, respectively) (p = 0.005).CONCLUSIONS:
The upper airway models closely simulated the in vivo deposition data. Optimizing the upper airway posture during inhalation via the nebulizers would be more efficient in increasing drug lung delivery than diluting their contents.
Buscar en Google
Banco de datos:
MEDLINE
Asunto principal:
Sistema Respiratorio
/
Nebulizadores y Vaporizadores
/
Imagen por Resonancia Magnética
/
Inhaladores de Dosis Medida
/
Albuterol
/
Pulmón
/
Modelos Biológicos
Tipo de estudio:
Prognostic_studies
Límite:
Humans
Idioma:
En
Año:
2005
Tipo del documento:
Article