Your browser doesn't support javascript.
loading
A simple and accurate method for calibrating the oscillation amplitude of tuning-fork based AFM sensors.
Liu, Jinquan; Callegari, Andrea; Stark, Martin; Chergui, Majed.
Afiliación
  • Liu J; Laboratoire de Spectroscopie Ultrarapide, Ecole Polytechnique Fédérale de Lausanne (EPFL), CH-1015 Lausanne, Switzerland.
Ultramicroscopy ; 109(1): 81-4, 2008 Dec.
Article en En | MEDLINE | ID: mdl-18947925
ABSTRACT
We have developed a simple and accurate method for calibrating the amplitude of vibration of quartz tuning fork sensors commonly used in atomic force- and near field optical-microscopy. Unlike interferometric methods, which require a complex optical setup, the method we present requires only a simple measurement of the electro-mechanical properties of the tuning-fork oscillator and can be performed in a matter of minutes without disturbing the experimental setup. Comparison with interferometric methods shows that an accuracy of better than few percent can be routinely achieved.

Texto completo: 1 Banco de datos: MEDLINE Idioma: En Año: 2008 Tipo del documento: Article

Texto completo: 1 Banco de datos: MEDLINE Idioma: En Año: 2008 Tipo del documento: Article