Sources of variability in measurements of cardiac troponin T in a community-based sample: the atherosclerosis risk in communities study.
Agarwal, Sunil K; Avery, Christy L; Ballantyne, Christie M; Catellier, Diane; Nambi, Vijay; Saunders, Justin; Sharrett, A Richey; Coresh, Josef; Heiss, Gerardo; Hoogeveen, Ron C.
Afiliación
  • Agarwal SK; Department of Epidemiology, University of North Carolina at Chapel Hill, Chapel Hill, NC, USA.
Clin Chem ; 57(6): 891-7, 2011 Jun.
Article en En | MEDLINE | ID: mdl-21519038
ABSTRACT

BACKGROUND:

Application of cardiac troponin T (cTnT) as a marker of myocyte damage requires knowledge of its measurement variability. Using a highly sensitive assay for measurement, we evaluated the long-term storage stability of plasma cTnT at -70 °C and the sources of cTnT variability.

METHODS:

Samples from the Atherosclerosis Risk in Communities study collected in 1996-1998 and 2005-2006 were assayed centrally to quantify variability in cTnT attributable to processing (replicates from same blood draw, n = 87), laboratory (replicates after freeze thaw, n = 29), short-term (n = 40) and long-term biological variation (repeat visit, n = 38), and degradation in frozen storage (n = 7677).

RESULTS:

Approximately 30% of this population-based cohort had cTnT concentrations below the detection limit (3 ng/L). Reliability coefficients for all paired comparisons exceeded 0.93 except for samples drawn 8 years apart (r = 0.36). Sources of cTnT variation (as CVs) were laboratory, 2.1% and 11.2% in those with and without heart failure, respectively; processing, 18.3%; biological, 16.6% at 6 weeks and 48.4% at 8 years. The reference change value at 6 weeks (68.5%) indicated that 4 samples are needed to determine a homeostatic set point within ±25%. The estimated cTnT degradation rate over the first year in long-term frozen storage was 0.36 ng/L per year.

CONCLUSIONS:

cTnT was detectable in approximately 70% of community-dwelling middle-aged study participants and stable in -70 °C storage. The variability in cTnT attributable to 1 freeze-thaw cycle is of small magnitude. The observed high laboratory and intraindividual (biological) reliability of cTnT support its use for population-based research, and in clinical settings that rely on classification and serial measurements.