Your browser doesn't support javascript.
loading
Pharmacokinetic and tolerability profiles of tobramycin nebuliser solution 300 mg/4 ml administered by PARI eFlow(®) rapid and PARI LC Plus(®) nebulisers in cystic fibrosis patients.
Govoni, Mirco; Poli, Gianluigi; Acerbi, Daniela; Santoro, Debora; Cicirello, Helen; Annoni, Ottavia; Ruzicka, Jozef.
Afiliación
  • Govoni M; Chiesi Farmaceutici S.p.A., Via Palermo, 26/A, 43100 Parma, Italy. m.govoni@chiesi.com
Pulm Pharmacol Ther ; 26(2): 249-55, 2013 Apr.
Article en En | MEDLINE | ID: mdl-23232039
ABSTRACT

BACKGROUND:

Tobramycin nebuliser solution (TNS) is indicated for maintenance therapy in cystic fibrosis (CF) patients with chronic Pseudomonas aeruginosa (PA) infections. Adherence to recommended therapy in CF has always been a challenge and new generation nebulisers are increasingly used "off label" to reduce the time required for inhalation, potentially improving patient compliance.

METHODS:

In this open-label, randomised, multi-centre, two-period crossover study, 27 CF patients with PA infection received TNS 300 mg/4 mL (TNS4) via the PARI eFlow(®) rapid or PARI LC Plus(®) nebuliser twice daily for 28 days in two study periods separated by a 4-week washout. The pharmacokinetic profile in plasma and sputum were determined after single and multiple dose administration on Day 1 and Day 28, respectively. Nebulisation times and general safety and tolerability profiles were evaluated throughout the study.

RESULTS:

Plasma tobramycin pharmacokinetic profiles were similar for the eFlow and LC Plus nebulisers both on Day 1 and Day 28. After multiple dose administration for 28 days, the eFlow/LC Plus ratio of geometric means for plasma C(max) and AUC(0-t), were 85.32 (90% CI, 61.24-118.86) and 87.44 (90% CI, 64.87-117.87), respectively. Despite the high variability, sputum tobramycin C(max) and AUC(0-t) for the eFlow on Day 28 tended to be higher than for the LC Plus (90% CI for the ratio, 86.11-226.45 and 81.81-236.71), respectively. Nebulisation times were significantly shorter for the eFlow with a median time for nebulisation of 5 min in comparison to 13 min for the LC Plus. Safety data confirmed a favourable safety profile for TNS4, with the majority of the findings being related to the underlying CF disease.

CONCLUSIONS:

Plasma and sputum pharmacokinetic data in CF patients with chronic PA infection support comparable pulmonary delivery and safety of TNS4 administered using different nebulisers, with a significantly shorter nebulisation time for the eFlow.
Asunto(s)

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Infecciones por Pseudomonas / Tobramicina / Fibrosis Quística / Antibacterianos Tipo de estudio: Clinical_trials Límite: Adolescent / Adult / Female / Humans / Male / Middle aged Idioma: En Año: 2013 Tipo del documento: Article

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Infecciones por Pseudomonas / Tobramicina / Fibrosis Quística / Antibacterianos Tipo de estudio: Clinical_trials Límite: Adolescent / Adult / Female / Humans / Male / Middle aged Idioma: En Año: 2013 Tipo del documento: Article