Observed Length Increases of Magnetically Controlled Growing Rods are Lower Than Programmed.
Gilday, Sarah E; Schwartz, Mark S; Bylski-Austrow, Donita I; Glos, David L; Schultz, Lindsay; O'Hara, Sara; Jain, Viral V; Sturm, Peter F.
Afiliación
  • Gilday SE; Cincinnati Children's Hospital Medical Center, Cincinnati, OH.
  • Schwartz MS; The Children's Hospital at Saint Francis, Warren Clinic, Tulsa, OK.
  • Bylski-Austrow DI; Cincinnati Children's Hospital Medical Center, Cincinnati, OH.
  • Glos DL; Cincinnati Children's Hospital Medical Center, Cincinnati, OH.
  • Schultz L; Cincinnati Children's Hospital Medical Center, Cincinnati, OH.
  • O'Hara S; Cincinnati Children's Hospital Medical Center, Cincinnati, OH.
  • Jain VV; Cincinnati Children's Hospital Medical Center, Cincinnati, OH.
  • Sturm PF; Cincinnati Children's Hospital Medical Center, Cincinnati, OH.
J Pediatr Orthop ; 38(3): e133-e137, 2018 Mar.
Article en En | MEDLINE | ID: mdl-29319661
ABSTRACT

BACKGROUND:

Magnetically controlled growing rods (MCGRs) are increasingly used in the treatment of early onset scoliosis (EOS). Few studies have reported whether desired lengthening can reliably be achieved, or if prior spine instrumentation and large tissue depths affect lengthening. In this clinical study of EOS patients, it was hypothesized that increases in rod length would equal programmed increases, patients with prior spine instrumentation would lengthen less than patients without prior surgery, and larger tissue depths would decrease lengthening success.

METHODS:

A retrospective chart review was conducted on EOS patients with single and dual MCGRs placed between April 2014 to September 2015 and distracted at a single institution. Rod distraction was measured at each visit using ultrasound. Differences between programmed and actual distraction for each patient, and between groups with and without prior spine instrumentation, were determined by 2-tailed t tests. Regression and correlation were used to determine the relationship between tissue depth and length increases.

RESULTS:

Thirty-one patients were included, 18 males, 13 females, age 8.1 (±2.5) years, with major curves measuring 60 (±14.6) degrees at time of MCGR insertion. In the 12 patients with prior instrumentation, time from initial growing rod placement to MCGR insertion was 23.1 (±10.6) months. The number of surgical procedures before MCGR insertion was 2.8 (±2.0). Total length increase relative to the programmed distraction was 86% (±21) (P<0.001). Length increases for patients with and without prior surgery were 87% (±23) and 86% (±19), respectively (P>0.9). Total lengthening was inversely proportional to tissue depth (r=0.38, P<0.01); the decrease in lengthening achieved was 2.1%/mm of tissue depth.

CONCLUSIONS:

Increases in rod length were 14% lower than the programmed distraction. Prior instrumentation did not impact the amount of rod distraction. Greater distance between the rod and the skin surface negatively affected the magnitude of distraction.
Asunto(s)

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Prótesis e Implantes / Escoliosis / Fijadores Internos Tipo de estudio: Diagnostic_studies / Observational_studies Límite: Child / Female / Humans / Male Idioma: En Año: 2018 Tipo del documento: Article

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Prótesis e Implantes / Escoliosis / Fijadores Internos Tipo de estudio: Diagnostic_studies / Observational_studies Límite: Child / Female / Humans / Male Idioma: En Año: 2018 Tipo del documento: Article